DQR ID | Subject | Data Streams Affected |
---|---|---|
D070412.1 | FKB/MET/M1 - Optical Rain Gauge Data Incorrect | fkbmetM1.b1 |
D070412.2 | FKB/MET/M1 - ORG Data incorrect during calibration | fkbmetM1.b1 |
D070417.1 | FKB/MET/M1 - Optical Rain Gauge Data Incorrect | fkbmetM1.b1 |
D080116.25 | FKB/MET/M1 - Optical Rain Gage Reporting Precip Constantly | fkbmetM1.b1 |
D081022.13 | HFE/MET/M1 - Wind Speed data contain offset due to induced voltage | hfemetM1.b1 |
D081022.9 | FKB/MET/M1 - Wind Speed data contain offset due to induced voltage | fkbmetM1.b1 |
D090216.3 | HFE/MET/M1 - Intermittent barometer failure | hfemetM1.b1 |
D090216.4 | HFE/MET/M1 - Intermittent PWD data | hfemetM1.b1 |
D090220.2 | HFE/MET/M1 - ORG Measuring Ice Crystals or Fog | hfemetM1.b1 |
D090220.3 | HFE/MET/M1 - Induced voltage on wind speed sensor cable | hfemetM1.b1 |
D090512.4 | GRW/MET/M1 - Tower lowered during installation | grwmetM1.b1 |
D091019.2 | GRW/MET/M1 - Temperature/RH Aspirator Failure | grwmetM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2007 | 0000 | 03/17/2007 | 0600 |
04/02/2007 | 0000 | 04/02/2007 | 0800 |
04/03/2007 | 0300 | 04/03/2007 | 0800 |
04/04/2007 | 2300 | 04/05/2007 | 0900 |
04/06/2007 | 0100 | 04/06/2007 | 0800 |
04/07/2007 | 0000 | 04/07/2007 | 0700 |
04/09/2007 | 0000 | 04/09/2007 | 0800 |
04/10/2007 | 0000 | 04/10/2007 | 0700 |
04/11/2007 | 0000 | 04/11/2007 | 0700 |
Subject: | FKB/MET/M1 - Optical Rain Gauge Data Incorrect |
DataStreams: | fkbmetM1.b1 |
Description: | The Optical Raingauge (ORG) data was incorrect during the listed times. Ice/frost was forming on the lenses causing the ORG to believe there was precipitation occurring. |
Measurements: | fkbmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2007 | 0736 | 04/08/2007 | 0745 |
Subject: | FKB/MET/M1 - ORG Data incorrect during calibration |
DataStreams: | fkbmetM1.b1 |
Description: | The ORG data was incorrect during calibration checks. The check mimics a precipiation event so the precip amounts that were collected were not actually measured. |
Measurements: | fkbmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2007 | 0000 | 04/28/2007 | 2359 |
Subject: | FKB/MET/M1 - Optical Rain Gauge Data Incorrect |
DataStreams: | fkbmetM1.b1 |
Description: | Precipitation was being reported on days with no rain. Problems with the grounding and the orientation of the receiver facing south (sunlight in the optics) caused erroneous readings. |
Measurements: | fkbmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/15/2007 | 0300 | 12/31/2007 | 2359 |
Subject: | FKB/MET/M1 - Optical Rain Gage Reporting Precip Constantly |
DataStreams: | fkbmetM1.b1 |
Description: | During the time period listed the ORG reported precip consistently. Reason unknown as deployment ended before resolution of problem. |
Measurements: | fkbmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2008 | 0000 | 12/28/2008 | 2159 |
Subject: | HFE/MET/M1 - Wind Speed data contain offset due to induced voltage |
DataStreams: | hfemetM1.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | hfemetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2007 | 0000 | 01/01/2008 | 2359 |
Subject: | FKB/MET/M1 - Wind Speed data contain offset due to induced voltage |
DataStreams: | fkbmetM1.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | fkbmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2008 | 0000 | 12/28/2008 | 2159 |
Subject: | HFE/MET/M1 - Intermittent barometer failure |
DataStreams: | hfemetM1.b1 |
Description: | Intermittent barometer failure occurred throughout the deployment. We were unable to pinpoint the cause of the failure. Typically one or two minutes of missing data occurred every few days. Since pressure data is typically invariant over 1 minute, linear interpolation should be fine to fill in missing data. |
Measurements: | hfemetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2008 | 0000 | 12/28/2008 | 2159 |
Subject: | HFE/MET/M1 - Intermittent PWD data |
DataStreams: | hfemetM1.b1 |
Description: | The Present Weather Detector had intermittent missing values throughout this deployment. Missing values occurred almost everyday, usually just one or two minutes per day. |
Measurements: | hfemetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/09/2008 | 1920 | 11/09/2008 | 1956 |
11/19/2008 | 2108 | 11/09/2008 | 2304 |
11/27/2008 | 1614 | 11/27/2008 | 2331 |
12/04/2008 | 1955 | 12/04/2008 | 2351 |
12/05/2008 | 0946 | 12/06/2008 | 0013 |
12/06/2008 | 1424 | 12/06/2008 | 2322 |
12/14/2008 | 1500 | 12/14/2008 | 1532 |
12/14/2008 | 1712 | 12/14/2008 | 1726 |
12/14/2008 | 1848 | 12/15/2008 | 0052 |
12/21/2008 | 0657 | 12/22/2008 | 0232 |
12/22/2008 | 0408 | 12/23/2008 | 0043 |
12/23/2008 | 1754 | 12/24/2008 | 0006 |
Subject: | HFE/MET/M1 - ORG Measuring Ice Crystals or Fog |
DataStreams: | hfemetM1.b1 |
Description: | The ORG was measuring consistent values for long periods of time. Typically this is indicative of fog, ice crystals in the air or poor performance of the lens heaters. |
Measurements: | hfemetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/2008 | 0000 | 12/28/2008 | 2359 |
Subject: | HFE/MET/M1 - Induced voltage on wind speed sensor cable |
DataStreams: | hfemetM1.b1 |
Description: | The cable for the wind monitor is not a twisted pair shielded cable. This caused an induced voltage on the wind sensor wires from the wind direction measurement excitation voltage. A small offset is added into the wind speeds. This value varies. Most frequently the values of 0.49, 0.588 and 0.686 m/s were added into the wind speed. On 11/27/2008 the value jumped to 0.98 m/s for most of the day. |
Measurements: | hfemetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2009 | 0000 | 04/20/2009 | 2359 |
Subject: | GRW/MET/M1 - Tower lowered during installation |
DataStreams: | grwmetM1.b1 |
Description: | Data was collected during set up and installation. Any data collected is suspect due to the tower being lowered and calibration checks occurring. The data should be viewed critically before being used. |
Measurements: | grwmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2009 | 0000 | 10/16/2009 | 0915 |
Subject: | GRW/MET/M1 - Temperature/RH Aspirator Failure |
DataStreams: | grwmetM1.b1 |
Description: | The aspirator fan on the solar shield failed causing incorrect readings on the temp, RH and calculated variables. |
Measurements: | grwmetM1.b1:
|