DQR ID | Subject | Data Streams Affected |
---|---|---|
D110628.2 | SGP/SIRS/C1 - NIP failure | sgpsirsC1.b1, sgpsirs20sC1.a0 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2011 | 0000 | 06/24/2011 | 1740 |
Subject: | SGP/SIRS/C1 - NIP failure |
DataStreams: | sgpsirsC1.b1, sgpsirs20sC1.a0 |
Description: | The short_direct_normal (NIP instrument) cable failed during this time period. Data are not recoverable. Users can derive an estimate of direct normal values using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) |
Measurements: | sgpsirs20sC1.a0:
sgpsirsC1.b1:
|