DQR ID | Subject | Data Streams Affected |
---|
D940822.1 | SGP/SMOS/E13 - Tower lowered for work on instrumentation | sgp1smosE13.a0, sgp30smosE13.a1 |
D941128.1 | | |
D950509.1 | SGP/SMOS/E13 - Relative Humidity and Vapor Pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1 |
D950509.2 | SGP/SMOS/E13 - Relative Humidity and Vapor Pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950509.3 | | |
D950612.1 | SGP/SMOS/E13 - Bad humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D950612.2 | SGP/SMOS/E13 - Bad humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D950612.3 | SGP/SMOS/E13 - Bad humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D950612.4 | SGP/SMOS/E13 - Bad humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D950707.1 | SGP/SMOS/E13 - Humidity and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950707.2 | SGP/SMOS/E13 - Humidity and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950707.3 | SGP/SMOS/E13 - Humidity and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950707.4 | SGP/SMOS/E13 - Humidity and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950808.2 | SGP/SMOS/E13 - Rain gauge and CR10 input wiring damaged by lightning | sgp1smosE13.a0, sgp30smosE13.a1 |
D950808.3 | SGP/SMOS/E13 - All data bad - internal battery voltage low | sgp1smosE13.a0, sgp30smosE13.a1 |
D950808.4 | SGP/SMOS/E13 - All RH and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D950808.5 | SGP/SMOS/E13 - All RH and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1 |
D950808.6 | SGP/SMOS/E13 - All RH data and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1 |
D950808.7 | SGP/SMOS/E13 - All RH and vapor pressure data bad | sgp1smosE13.a0, sgp30smosE13.a1 |
D950828.4 | SGP/SMOS/E11 - Bad Humidity Data | sgp1smosE11.a0, sgp30smosE11.a1, sgp1440smosE11.a0 |
D950828.5 | SGP/SMOS/E13 - Tower lowered for maintenance | sgp1smosE13.a0, sgp30smosE13.a1 |
D951009.10 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D951009.6 | | |
D951009.7 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D951009.9 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D951101.4 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1 |
D951114.3 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D951114.4 | SGP/SMOS/E13 - Incorrect humidity data | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D960311.1 | SGP/SMOS/E13 - Error in Snow Depth Output | sgp30smosE13.a1, sgp1440smosE13.a0 |
D960523.12 | SGP/SMOS/E13 - Tower down for maintenance | sgp1smosE13.a0, sgp30smosE13.a1 |
D960523.13 | | |
D961210.15 | | |
D970214.2 | SGP/SMOS/E13 - Questionable data | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D970325.5 | SGP/SMOS/E13 - T/RH Probe Malfunction | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D970325.6 | SGP/SMOS/E13 - T/RH Probe Malfunction | sgp1smosE13.a0, sgp30smosE13.a1 |
D970325.7 | SGP/SMOS/E13 - T/RH Probe Malfunction | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D970721.3 | | |
D971112.13 | | |
D971215.1 | SGP/SMOS/E13 - Questionable E13 SMOS humidity data APR - NOV 95 | sgp1smosE13.00, sgp30smosE13.a1, sgp1440smosE13.a0 |
D980601.11 | | |
D990512.10 | | |
D991109.11 | SG/SMOS/E13 - E13 SMOS down for tower replacement | sgpsmosE13.01, sgp1smosE13.00, sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D991109.20 | SGP/SMOS/E13 - E13 SMOS lowered to surface to install counterweights | sgpsmosE13.01, sgp1smosE13.00, sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
Subject: | SGP/SMOS/E13 - All data bad - internal battery voltage low |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1
|
Description: | All data incorrect. Internal battery voltage low, apparently not charging. |
Measurements: | sgp30smosE13.a1: - Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- Snow, depth, 30-min intervals(snow)
- vbat
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Wind speed, 30-min intervals, standard deviation(sd_wspd)
- lat
- Humidity, relative(rh)
- Unit Vector Wind Direction(wdir)
- base_time
- lon
- Temperature, air(temp)
- time_offset
- alt
- snow_sen
- sd_vap_pres
- Mean Wind Speed(wspd)
- sd_bar_pres
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
- Precipitation, total(precip)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
sgp1smosE13.a0: - Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- Unit Vector Wind Direction(wdir)
- alt
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Precipitation, total(precip)
- Temperature, air(temp)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
- lat
- Mean Wind Speed(wspd)
- time_offset
- lon
- Humidity, relative(rh)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- base_time
|
Subject: | SGP/SMOS/E13 - All RH and vapor pressure data bad |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | All Relative Humidity and Vapor Pressure data incorrect. Sensor may have gotten
wet. Since rain gauge wasn't working, it cannot be verified. |
Measurements: | sgp1440smosE13.a0: - Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Time of Maximum Relative Humidity(time_max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
sgp1smosE13.a0: - Humidity, relative(rh)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
sgp30smosE13.a1: - Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- sd_vap_pres
- Humidity, relative(rh)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
|
Subject: | SGP/SMOS/E13 - Tower lowered for maintenance |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1
|
Description: | Tower lowered for SIROS maintenance. |
Measurements: | sgp30smosE13.a1: - Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- Snow, depth, 30-min intervals(snow)
- vbat
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Wind speed, 30-min intervals, standard deviation(sd_wspd)
- lat
- Humidity, relative(rh)
- Unit Vector Wind Direction(wdir)
- base_time
- lon
- Temperature, air(temp)
- time_offset
- alt
- snow_sen
- sd_vap_pres
- Mean Wind Speed(wspd)
- sd_bar_pres
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
- Precipitation, total(precip)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
sgp1smosE13.a0: - Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- Unit Vector Wind Direction(wdir)
- alt
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Precipitation, total(precip)
- Temperature, air(temp)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
- lat
- Mean Wind Speed(wspd)
- time_offset
- lon
- Humidity, relative(rh)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- base_time
|
Subject: | SGP/SMOS/E13 - Tower down for maintenance |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1
|
Description: | SMOS E13 tower was down for maintenance on 13 APR 96 between 1530
and 1605 GMT. All wind, temperature, and humidity data taken during
this period are incorrect. Rainfall and Barometric Pressure data are OK. |
Measurements: | sgp30smosE13.a1: - Temperature, air(temp)
- Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Wind speed, 30-min intervals, standard deviation(sd_wspd)
- sd_vap_pres
- Humidity, relative(rh)
- Unit Vector Wind Direction(wdir)
- Mean Wind Speed(wspd)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
sgp1smosE13.a0: - Wind direction, 30-min intervals, standard deviation(sd_deg)
- Unit Vector Wind Direction(wdir)
- Mean Wind Speed(wspd)
- Humidity, relative(rh)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- Temperature, air(temp)
|
Subject: | SGP/SMOS/E13 - Questionable data |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | The values of relative humidity reported by the SMOS at E13 between
18 NOV 96 and 1930 GMT on 30 JAN 97 are too high. Saturation values
of 104% to 105% have been reported during this period. A 6-month
Verification Check was made on 18 NOV 96 and the SMOS probe did not
agree with the reference at that time. The SMOS probe was finally
replaced on 30 JAN 97. Data reported since that time appears reasonable. |
Measurements: | sgp1440smosE13.a0: - Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Time of Maximum Relative Humidity(time_max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
sgp1smosE13.a0: - Humidity, relative(rh)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
sgp30smosE13.a1: - Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- sd_vap_pres
- Humidity, relative(rh)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
|
Subject: | SGP/SMOS/E13 - T/RH Probe Malfunction |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | When the SMOS E13 T/RH probe was replaced on 30 JAN 97, it
was not adequately sealed to prevent rain infiltration. During a
rain, which fell between 20 FEB 97 at 0730 GMT and 22 FEB 97 at 1230
GMT, the RH probe malfunctioned apparently due to becoming wet. |
Measurements: | sgp1440smosE13.a0: - Time of Maximum Relative Humidity(time_max_rh)
- Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Temperature, air, at 60-m height, 24-hr maximum(max_temp)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Time of Minimum Temperature(time_min_temp)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
- Time of Maximum Temperature(time_max_temp)
- Temperature, air, at 60-m height, 24-hr minimum(min_temp)
sgp1smosE13.a0: - Humidity, relative(rh)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Temperature, air(temp)
sgp30smosE13.a1: - Temperature, air(temp)
- Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- sd_vap_pres
- Humidity, relative(rh)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
|
Subject: | SGP/SMOS/E13 - T/RH Probe Malfunction |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | When the SMOS E13 T/RH probe was replaced on 30 JAN 97, it
was not adequately sealed to prevent rain infiltration. During a
rain, which fell on 27 FEB 97 between 0230 and 1630 GMT, the RH probe
malfunctioned apparently due to becoming wet. |
Measurements: | sgp1440smosE13.a0: - Time of Maximum Relative Humidity(time_max_rh)
- Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Temperature, air, at 60-m height, 24-hr maximum(max_temp)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Time of Minimum Temperature(time_min_temp)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
- Time of Maximum Temperature(time_max_temp)
- Temperature, air, at 60-m height, 24-hr minimum(min_temp)
sgp1smosE13.a0: - Humidity, relative(rh)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Temperature, air(temp)
sgp30smosE13.a1: - Temperature, air(temp)
- Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- sd_vap_pres
- Humidity, relative(rh)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
|
Subject: | SGP/SMOS/E13 - Questionable E13 SMOS humidity data APR - NOV 95 |
DataStreams: | sgp1smosE13.00, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | When the E13 SMOS naturally aspirated radiation shield was replaced with a fan aspirated
radiation shield on 14 APR 95 at 2030 GMT, the new shield was not adequately sealed to
prevent penetration of water, whether from rain or from dew, from entering the radiation
shield and causing incorrect humidity data.
After numerous attempts, the radiation shield was successfully sealed early in NOV 95.
Separate DQR's have been issued for humidity data that was obviously erroneous, but on
numerous occasions during that period the humidity data has looked suspicious (e.g., see PIF
P960624.1).
Therefore, all humidity data reported by the E13 SMOS during that period should be
considered questionable. |
Measurements: | sgp1smosE13.00: - Raw data stream - documentation not supported
sgp1440smosE13.a0: - Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Time of Maximum Relative Humidity(time_max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
sgp30smosE13.a1: - Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- sd_vap_pres
- Humidity, relative(rh)
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
|
Subject: | SG/SMOS/E13 - E13 SMOS down for tower replacement |
DataStreams: | sgpsmosE13.01, sgp1smosE13.00, sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | The E13 SMOS was shut down from 14 APR 1999 at 2100 GMT until 3 MAY 1999 at 2055 GMT
while the tower was replaced. |
Measurements: | sgp1smosE13.00: - Raw data stream - documentation not supported
sgp1440smosE13.a0: - Minimum Barometric Pressure(min_bar_pres)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Time of Maximum Snow Depth(time_max_snow)
- Temperature, air, at 60-m height, 24-hr maximum(max_temp)
- Maximum Barometric Pressure(max_bar_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Highest gust within hour is reported(max_wspd)
- Time of Minimum Temperature(time_min_temp)
- Minimum Wind Speed(min_wspd)
- Time of Minimum Barometric Pressure(time_min_bar_pres)
- base_time
- alt
- Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- lon
- Maximum Snow Depth(max_snow)
- Time of Minimum Wind Speed(time_min_wspd)
- Time of Maximum Barometric Pressure(time_max_bar_pres)
- Time of Minimum Snow Depth(time_min_snow)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Minimum Snow Depth(min_snow_depth)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
- Temperature, air, at 60-m height, 24-hr minimum(min_temp)
- Time of Maximum Relative Humidity(time_max_rh)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Time of Maximum Wind Speed(time_max_wspd)
- Time of Relative Humidity(time_min_rh)
- lat
- Time of Maximum Temperature(time_max_temp)
- time_offset
sgpsmosE13.01: - Raw data stream - documentation not supported
sgp30smosE13.a1: - Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- vbat
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- lat
- lon
- alt
- time_offset
- sd_vap_pres
- sd_bar_pres
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
- Precipitation, total(precip)
- Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Snow, depth, 30-min intervals(snow)
- Wind speed, 30-min intervals, standard deviation(sd_wspd)
- Humidity, relative(rh)
- Unit Vector Wind Direction(wdir)
- base_time
- Temperature, air(temp)
- snow_sen
- Mean Wind Speed(wspd)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
sgp1smosE13.a0: - alt
- Mean Wind Speed(wspd)
- time_offset
- Humidity, relative(rh)
- Wind speed, vector-averaged, at 10-m height, 30-min intervals(wspd_va)
- Pressure, atmospheric, at 1-m height, 30-min intervals(bar_pres)
- Unit Vector Wind Direction(wdir)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Precipitation, total(precip)
- Temperature, air(temp)
- Wind direction, 30-min intervals, standard deviation(sd_deg)
- lat
- lon
- base_time
|
Subject: | SGP/SMOS/E13 - E13 SMOS lowered to surface to install counterweights |
DataStreams: | sgpsmosE13.01, sgp1smosE13.00, sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0
|
Description: | The E13 SMOS was lowered to the surface on 24 SEP 1999 between 1300 and 1320 GMT
while counterweights were installed. All temperature and humidity data are questionable
during this period. |
Measurements: | sgp1smosE13.00: - Raw data stream - documentation not supported
sgp1440smosE13.a0: - Time of Maximum Relative Humidity(time_max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Temperature, air, at 60-m height, 24-hr maximum(max_temp)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Humidity, relative, at 60-m height, 24-hr minimum(min_rh)
- Time of Relative Humidity(time_min_rh)
- Time of Minimum Temperature(time_min_temp)
- Time of Maximum Temperature(time_max_temp)
- Humidity, relative, at 60-m height, 24-hr maximum(max_rh)
- Pressure, vapor, at 60-m height, 24-hr minimum(min_vap_pres)
- Pressure, vapor, at 60-m height, 24-hr maximum(max_vap_pres)
- Temperature, air, at 60-m height, 24-hr minimum(min_temp)
sgpsmosE13.01: - Raw data stream - documentation not supported
sgp30smosE13.a1: - Temperature, air, 30-min intervals, standard deviation(sd_temp)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Humidity, relative(rh)
- Temperature, air(temp)
- sd_vap_pres
- Humidity, relative, 30-min intervals, standard deviation(sd_rh)
sgp1smosE13.a0: - Unit Vector Wind Direction(wdir)
- Pressure, vapor, 60-m height, 30-min intervals(vap_pres)
- Temperature, air(temp)
- Mean Wind Speed(wspd)
- Humidity, relative(rh)