DQR ID | Subject | Data Streams Affected |
---|---|---|
D111006.1 | GAN/MWR/M1 - Missing data | ganmwrlosM1.b1, ganmwrtipM1.a1 |
D111006.10 | GAN/MET/M1 - Tower lowered and tipped for maintenance | ganmetM1.b1 |
D111006.9 | GAN/MET/M1 - Met data questionable | ganmetM1.b1 |
D111017.1 | GAN/GNDRAD/M1 - IRT calibration error | gangndrad60sM1.b1 |
D111025.1 | GAN/MET/M1 - Instrument problem | ganmetM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2011 | 0000 | 10/05/2011 | 2118 |
Subject: | GAN/MWR/M1 - Missing data |
DataStreams: | ganmwrlosM1.b1, ganmwrtipM1.a1 |
Description: | The MWR was unable to communicate with the software. The problem was identified as an incorrect port configuration. |
Measurements: | ganmwrlosM1.b1:
ganmwrtipM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2011 | 0348 | 10/05/2011 | 0444 |
Subject: | GAN/MET/M1 - Tower lowered and tipped for maintenance |
DataStreams: | ganmetM1.b1 |
Description: | The met tower was lowered for a PIR replacement. All met tower data are questionable during this period. |
Measurements: | ganmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2011 | 1030 | 09/30/2011 | 2359 |
Subject: | GAN/MET/M1 - Met data questionable |
DataStreams: | ganmetM1.b1 |
Description: | Met data are questionable during this time period as the system was being deployed and in various states of operation. |
Measurements: | ganmetM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2011 | 0742 | 10/09/2011 | 0413 |
Subject: | GAN/GNDRAD/M1 - IRT calibration error |
DataStreams: | gangndrad60sM1.b1 |
Description: | The IRT surface temperature reported was incorrect because the analog output configuration did not agree with the data logger scale factors. The IRT configuration was corrected on 10/9. |
Measurements: | gangndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/04/2011 | 0410 | 10/09/2011 | 1200 |
Subject: | GAN/MET/M1 - Instrument problem |
DataStreams: | ganmetM1.b1 |
Description: | Intermittent backscatter alarms were tripping on the PWD. This suggested something is blocking (dew) or getting into the measurement path (spider or ants). There was no clear indication of the problem. |
Measurements: | ganmetM1.b1:
|