DQR ID | Subject | Data Streams Affected |
---|---|---|
D040416.1 | TWP/GNDRAD/C3 - IRT bias | twpgndrad60sC3.b1 |
D040812.5 | TWP/SKYRAD/C3 - UVB Data questionable/incorrect due to uncontrolled sensor temperature | twpskyrad20sC3.a1, twpskyrad60sC3.b1 |
D040904.2 | TWP/SKYRAD/C3 - Incorrect NIP cal coefficient | twpskyrad60sC3.b1 |
D050104.1 | TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect | twpskyradC3.00, twpskyrad60sC3.b1 |
D050221.2 | TWP/GNDRAD/C3 - IRT bias | twpgndrad60sC3.b1 |
D050221.3 | TWP/GNDRAD/C3 - Reprocess: IRT with wrong calibration | twpgndrad60sC3.b1 |
D051202.3 | TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 downwelling longwave | twpskyrad60sC3.b1 |
D060630.33 | TWP/SKYRAD/C3 - Reprocess: Longwave Calibration error | twpskyrad60sC3.b1 |
D060630.36 | TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error | twpgndrad60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/07/2004 | 0230 | 04/08/2004 | 1200 |
04/12/2004 | 0600 | 04/27/2004 | 1400 |
05/08/2004 | 0300 | 05/15/2004 | 2200 |
05/27/2004 | 1100 | 06/09/2004 | 0000 |
12/29/2004 | 1100 | 01/06/2005 | 2300 |
Subject: | TWP/GNDRAD/C3 - IRT bias |
DataStreams: | twpgndrad60sC3.b1 |
Description: | The IRT behaved erratically, jumping frequently from reporting about 50 Wm^(-2) higher than the PIR for surface irradiance values to about 100Wm^(-2) lower than the PIR. Generally, the IRT measured lower surface temperatures than expected from the PIR. |
Measurements: | twpgndrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2003 | 0000 | 05/19/2004 | 2200 |
Subject: | TWP/SKYRAD/C3 - UVB Data questionable/incorrect due to uncontrolled sensor temperature |
DataStreams: | twpskyrad20sC3.a1, twpskyrad60sC3.b1 |
Description: | The UVB was not temperature controlled to 25C as required to ensure accurate data. UVB data should ONLY be used after checking both the bit_data and twpskyrad20sC1.a1 uvb_temp or inst_down_uvb_temp. UVB data where the UVB temperature was outside the range of 24-26C should not be used. |
Measurements: | twpskyrad60sC3.b1:
twpskyrad20sC3.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/24/2004 | 0000 | 09/02/2004 | 2350 |
Subject: | TWP/SKYRAD/C3 - Incorrect NIP cal coefficient |
DataStreams: | twpskyrad60sC3.b1 |
Description: | During our site visit comparison at Darwin we observed that the skyrad NIP was reading about 50 w/m2 lower than the comparison cavity radfiometer and the replacement NIP on the comparison stand. We found that the calibration coefficient for the NIP (responsivity at 45o)was put in as 8.64X10-6. The NIP was new and had not been calibrated at SGP. The factory number was 8.04X10-6. We corrected the calibration coefficient and the instruments agreed to better than 5 w/m2. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2004 | 0000 | 09/02/2004 | 0400 |
Subject: | TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect |
DataStreams: | twpskyradC3.00, twpskyrad60sC3.b1 |
Description: | On 05/20 during installtion of the new Campbell datalogger system and associated program the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient. A coefficient of 116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV. TWP reports SV0405D and SV0409D detail the activities which lead to this DQR. |
Measurements: | twpskyrad60sC3.b1:
twpskyradC3.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/2004 | 0700 | 12/25/2004 | 0300 |
Subject: | TWP/GNDRAD/C3 - IRT bias |
DataStreams: | twpgndrad60sC3.b1 |
Description: | The IRT measured higher surface temperatures than expected from the PIR, even at night. |
Measurements: | twpgndrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/17/2004 | 0000 | 11/19/2004 | 0105 |
Subject: | TWP/GNDRAD/C3 - Reprocess: IRT with wrong calibration |
DataStreams: | twpgndrad60sC3.b1 |
Description: | A spare IRT was installed that was not configured properly. |
Measurements: | twpgndrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/2004 | 0000 | 09/09/2005 | 2012 |
Subject: | TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were installed. The result of no delay in these measurements results in an error in the measured resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 irradiances. The errors are inversely proportial to temperature. Users can apply temperature dependent irradiance corrections as shown in the table below. Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 W/m2. The error corrections become the same order of magnitude or larger than the instrument uncertainties at or below 260K or 9F. Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance DIR1 DIR1 DIR2 DIR2 T[F] Tcase[K] Correction[W/m2] Tdome[K] Correction[W/m2] 109.4 316.13 0.00 316.13 0.00 100.5 311.20 -0.15 311.20 0.12 93.2 307.14 -0.12 307.14 0.10 87.0 303.70 -0.10 303.70 0.08 81.6 300.70 -0.17 300.70 0.14 77.0 298.15 -0.22 298.15 0.18 59.4 288.35 -0.34 288.35 0.27 47.4 281.72 -0.23 281.72 0.18 38.5 276.74 -0.42 276.74 0.34 31.3 272.77 -0.48 272.77 0.38 25.4 269.48 -0.61 269.48 0.49 20.3 266.67 -0.67 266.67 0.54 16.0 264.25 -0.82 264.25 0.66 12.1 262.11 -0.98 262.11 0.79 8.7 260.19 -0.97 260.19 0.78 5.6 258.46 -1.07 258.46 0.86 2.7 256.89 -1.17 256.89 0.94 0.1 255.44 -1.33 255.44 1.07 -2.3 254.11 -1.41 254.11 1.13 -4.5 252.87 -1.43 252.87 1.15 -6.6 251.71 -1.52 251.71 1.22 -8.5 250.63 -1.63 250.63 1.31 -10.4 249.62 -1.73 249.62 1.39 -12.2 248.58 -1.79 248.58 1.44 -15.4 246.82 -1.93 246.82 1.55 -18.3 245.23 -2.05 245.23 1.65 -20.9 243.78 -2.22 243.78 1.78 -23.3 242.44 -2.36 242.44 1.89 -25.5 241.21 -2.58 241.21 2.07 -27.5 240.07 -2.65 240.07 2.13 -29.5 239.01 -2.75 239.01 2.21 -31.3 238.00 -3.02 238.00 2.42 |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2004 | 2200 | 12/21/2005 | 2136 |
Subject: | TWP/SKYRAD/C3 - Reprocess: Longwave Calibration error |
DataStreams: | twpskyrad60sC3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2004 | 0100 | 12/21/2005 | 2137 |
Subject: | TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC3.b1:
|