DQR ID | Subject | Data Streams Affected |
---|---|---|
D120420.10 | SGP/QCRAD/E11 - Upwelling LW flux appears bad | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
D120509.3 | SGP/QCRAD/E11 - shortwave fluxes too low | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
D120510.1 | SGP/QCRAD/E11 - Instrument multiplexer failure | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/11/1999 | 2100 | 05/19/1999 | 1900 |
Subject: | SGP/QCRAD/E11 - Upwelling LW flux appears bad |
DataStreams: | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
Description: | Longwave upwelling flux values appear incorrect during this time. The cause is not known. Currently there are no instrument DQRs describing this problem. |
Measurements: | sgpqcrad1longE11.c1:
sgpqcrad1longE11.c2:
sgpqcrad1longE11.s1:
sgpqcrad1longE11.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/1995 | 0000 | 09/25/1995 | 2200 |
Subject: | SGP/QCRAD/E11 - shortwave fluxes too low |
DataStreams: | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
Description: | Shortwave fluxes are too low during this time (including best estimate, global, direct normal, diffuse, upwelling shortwave). The cause of this problem is not known. No instrument DQRs available describing this problem. |
Measurements: | sgpqcrad1longE11.c1:
sgpqcrad1longE11.c2:
sgpqcrad1longE11.s1:
sgpqcrad1longE11.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/2008 | 1300 | 11/10/2008 | 2359 |
Subject: | SGP/QCRAD/E11 - Instrument multiplexer failure |
DataStreams: | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s1, sgpqcrad1longE11.s2 |
Description: | The multiplexer failed intermittently during this time. The down_short_hemisp, up_long_hemisp and down_long_hemisp were very noisy. The SIRS multiplexer wiring or mux itself was failing during this time. Ops replaced the mux on 12/2. The problem is documented in instrument DQR D081029.2 |
Measurements: | sgpqcrad1longE11.c1:
sgpqcrad1longE11.c2:
sgpqcrad1longE11.s1:
sgpqcrad1longE11.s2:
|