DQR ID | Subject | Data Streams Affected |
---|---|---|
D040526.6 | SGP/ECOR/E3 - Missing data | sgp30ecorE3.b1 |
D040617.2 | SGP/ECOR/E1 - Missing data | sgp30ecorE1.b1 |
D040804.1 | SGP/ECOR/E3 - Failed IRGA | sgp30ecorE3.b1 |
D041025.2 | SGP/ECOR/E1 - System time drift | sgp30ecorE1.b1, sgpecorE1.00 |
D041221.2 | SGP/ECOR/E3 - Data transfer problem | sgp30ecorE3.b1, sgpecorE3.00 |
D051101.6 | SGP/ECOR/E1 - ECOR Pressure and Temperature Missing | sgp30ecorE1.b1 |
D051101.7 | SGP/ECOR/E1 - CO2 and H2O Fluxes Incorrect | sgp30ecorE1.b1 |
D051101.8 | SGP/ECOR/E3 - CO2/H2O Sensor Not Working | sgp30ecorE3.b1 |
D090902.9 | SGP/ECOR/E3 - Pressure and Temperature Missing | sgp30ecorE3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/25/2004 | 0300 | 06/16/2004 | 1730 |
07/01/2005 | 0200 | 07/15/2005 | 1600 |
07/19/2005 | 1300 | 07/27/2005 | 1500 |
06/07/2006 | 0030 | 06/14/2006 | 1600 |
08/03/2006 | 0130 | 08/09/2006 | 1700 |
Subject: | SGP/ECOR/E3 - Missing data |
DataStreams: | sgp30ecorE3.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp30ecorE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/2004 | 0900 | 09/10/2004 | 1800 |
10/03/2004 | 1500 | 10/05/2004 | 1830 |
05/24/2005 | 0800 | 06/07/2005 | 1800 |
09/06/2005 | 1900 | 09/20/2005 | 2030 |
05/31/2006 | 2000 | 08/08/2006 | 2230 |
Subject: | SGP/ECOR/E1 - Missing data |
DataStreams: | sgp30ecorE1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2004 | 1730 | 06/30/2004 | 1730 |
Subject: | SGP/ECOR/E3 - Failed IRGA |
DataStreams: | sgp30ecorE3.b1 |
Description: | Infrared gas analyzer (IRGA) failed 6/16; a new IRGA was installed on 6/30. All data related to CO2 and water vapor concentration measurements (including CO2 flux and latent heat flux) are incorrect for the period of June 16, 2004, 17:30 to June 30, 2004, 17:30. Wind and sensible heat flux data are OK. |
Measurements: | sgp30ecorE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2004 | 2030 | 10/25/2004 | 1735 |
Subject: | SGP/ECOR/E1 - System time drift |
DataStreams: | sgp30ecorE1.b1, sgpecorE1.00 |
Description: | Network Time Protocol (NTP) client was not working at the ECOR computer resulting in an ECOR system time drift of about 6s per hour. Time adjustment by +15 minutes occurred at 17:35:25, Oct. 25, 2004. |
Measurements: | sgp30ecorE1.b1:
sgpecorE1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/19/2004 | 1353 | 12/20/2004 | 0931 |
Subject: | SGP/ECOR/E3 - Data transfer problem |
DataStreams: | sgp30ecorE3.b1, sgpecorE3.00 |
Description: | Serial communication with one sensor (the IRGA) were partially disrupted due to deteriorating hardware from 13:53, Dec. 19, 2004 to 09:31 Dec. 20, 2004. Serial data from the IRGA is mainly used to get sensor status codes, so this might lead to false increase of "bad" point count. It seems that IRGA analog outputs (stored in raw sonic data file, primary source of data) were not affected, so the flux data may be reprocessed or verified using raw sonic data files and ignoring IRGA sensor status. |
Measurements: | sgpecorE3.00:
sgp30ecorE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2005 | 1400 | 05/31/2006 | 1930 |
Subject: | SGP/ECOR/E1 - ECOR Pressure and Temperature Missing |
DataStreams: | sgp30ecorE1.b1 |
Description: | The pressure and temperature from the CO2/H2O sensor is missing due to a failure of some kind. |
Measurements: | sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/18/2005 | 2000 | 10/21/2005 | 1800 |
Subject: | SGP/ECOR/E1 - CO2 and H2O Fluxes Incorrect |
DataStreams: | sgp30ecorE1.b1 |
Description: | The CO2 and H2O fluxes (lv_e) were incorrect because of a LI-7500 failure. |
Measurements: | sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2005 | 1100 | 02/08/2006 | 1700 |
Subject: | SGP/ECOR/E3 - CO2/H2O Sensor Not Working |
DataStreams: | sgp30ecorE3.b1 |
Description: | The CO2/H2O sensor failed. Means and fluxes of CO2 and H2O (lv_e) are incorrect or missing. |
Measurements: | sgp30ecorE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/2009 | 1600 | 10/24/2009 | 2100 |
Subject: | SGP/ECOR/E3 - Pressure and Temperature Missing |
DataStreams: | sgp30ecorE3.b1 |
Description: | Pressure and temperature from the LI-7500 are missing. This did not affect the sensible and latent heat fluxes, however, much of the ECOR data is missing during this period. Attempts to correct the problem were unsuccessful until the time of site decommissioning. This facility was decommissioned and maintenance efforts ceased 10/1/2009. |
Measurements: | sgp30ecorE3.b1:
|