Description: | Tracker appears to be out of alignment or inoperative during these times. The
short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data appear to be bad and not
recoverable.
Several of these periods are documented in instrument DQRs:
D110831.1, D090328.1, D081222.3, D040107.1, D011207.1, D011019.8 |
Measurements: | twpqcrad1longC2.c2: - Best Estimate Global Downwelling Shortwave Hemispheric Irradiance(BestEstimate_down_short_hemisp)
- Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
- Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
twpqcrad1longC2.c1: - Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
- Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
- Best Estimate Global Downwelling Shortwave Hemispheric Irradiance(BestEstimate_down_short_hemisp)
twpqcrad1longC2.s1: - Best Estimate Global Downwelling Shortwave Hemispheric Irradiance(BestEstimate_down_short_hemisp)
- Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
- Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
twpqcrad1longC2.s2: - Radiation, shortwave, direct normal irradiance, 1-min avg(short_direct_normal)
- Best Estimate Global Downwelling Shortwave Hemispheric Irradiance(BestEstimate_down_short_hemisp)
- Radiation, shortwave, downwelling diffuse hemispheric irradiance, 1-min avg(down_short_diffuse_hemisp)
|