DQR ID | Subject | Data Streams Affected |
---|---|---|
D060928.3 | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry | sgpsondewnpnC1.b1 |
D130108.5 | SGP/SIRS/E34 - Tracker failure | sgpsirs20sE34.a0, sgpsirsE34.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2006 | 1728 | 12/31/9999 | 2359 |
Subject: | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | The output precision of the following variables was increased 20060928 alt changed from 1m to 0.1m dp changed from 0.1degC to 0.01degC pres changed from 0.1 hPa to 0.01 hPa RH changed from 1%RH to 0.01%RH tdry changed from .1degC to 0.01degC Note that the increase in precision does not necessarily imply an increase in accuracy. Users are cautioned to continue to use their best scientific judgement when using these or any other data. |
Measurements: | sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/26/2012 | 0000 | 01/03/2013 | 1923 |
Subject: | SGP/SIRS/E34 - Tracker failure |
DataStreams: | sgpsirs20sE34.a0, sgpsirsE34.b1 |
Description: | The tracker at failed to wake up on 12/26/2012. The problem persisted throughout the time range specified. On 1/3/2013 the tracker was power cycled and returned to normal operation. The tracker malfunction adversely affected the data from the direct normal instrument (short_direct_normal). Also affected is the diffuse shaded global data (down_short_diffuse). |
Measurements: | sgpsirsE34.b1:
sgpsirs20sE34.a0:
|