DQR ID | Subject | Data Streams Affected |
---|---|---|
D140121.3 | TWP/RL/C3 - Missing T1 and/or T2 (rotational Raman) channels | twprlprofmerge1turnC3.c0 |
D140223.1 | TWP/RL/C3 - Invalid depolarization channel data | twprlprofmerge1turnC3.c0 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2014 | 0000 | 02/11/2014 | 0000 |
12/28/2013 | 0000 | 01/09/2014 | 0059 |
Subject: | TWP/RL/C3 - Missing T1 and/or T2 (rotational Raman) channels |
DataStreams: | twprlprofmerge1turnC3.c0 |
Description: | Data for the T1 and/or T2 (rotation Raman) channels are missing due to failures of multiple high voltage power supplies for PMTs. As a result, it is not be possible to derive temperature profiles from the Raman lidar during the effected period. Temperature profile data from the Raman lidar are incorrect. Do not use for analysis. |
Measurements: | twprlprofmerge1turnC3.c0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2014 | 1800 | 01/15/2014 | 2300 |
Subject: | TWP/RL/C3 - Invalid depolarization channel data |
DataStreams: | twprlprofmerge1turnC3.c0 |
Description: | The depolarization channel was not working properly during the period in question due to a failure of the high voltage power supply for the PMT. |
Measurements: | twprlprofmerge1turnC3.c0:
|