Description: | On March 27 2011 at 0100 GMT the Raman lidar's instrument computer automatically advanced
it's time by one hour (transition to daylight savings time). The problem was eventually
fixed on March 29 2011 at 1300 when the clock was set back one hour to the correct time.
The ingest experienced problems when the clock was set back one hour.
The time arrays in the following files are set one hour ahead of what they should be:
1. sgprlC1.a0.20110327.000001.cdf for profile indices equal to or greater than 355
2. all of sgprlC1.a0.20110327.120001.cdf
3. all of sgprlC1.a0.20110328.134549.cdf
4. all of sgprlC1.a0.20110329.000005.cdf
5. sgprlC1.a0.20110329.120005.cdf for profile indices less than or equal to 72
Also, the time arrays in sgprlC1.a0.20110329.120005.cdf between profile indices 73 through
641 are incorrect. In this case, the offset varies approximately linearly with profile
index from 1 hour at index 73 to 0 at index 641. The ingest apparently introduced some
unphysical variations in time in this part of the file. |