DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120414.6 | SGP/QCRAD/E16 - Tracker alignment issue | sgpqcrad1longE16.s2 | Incorrect |
D120510.28 | SGP/QCRAD/E16 - Tc/Td bad or noisy | sgpqcrad1longE16.s2 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/14/1999 | 1500 | 03/24/1999 | 2359 | Incorrect |
10/05/1999 | 0600 | 10/20/1999 | 1400 | Incorrect |
Subject: | SGP/QCRAD/E16 - Tracker alignment issue |
DataStreams: | sgpqcrad1longE16.s2 |
Description: | Tracker appears to be out of alignment or inoperative during these times. The short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data appear to be bad and not recoverable. There were no instrument DQRs on this problem at the time this report was submitted. |
Measurements: | sgpqcrad1longE16.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/18/2007 | 1900 | 01/16/2008 | 1800 | Suspect |
05/01/1990 | 0000 | 08/01/1999 | 0000 | Suspect |
03/24/2003 | 1500 | 04/15/2003 | 0300 | Suspect |
04/03/1997 | 1500 | 05/07/1997 | 1700 | Suspect |
Subject: | SGP/QCRAD/E16 - Tc/Td bad or noisy |
DataStreams: | sgpqcrad1longE16.s2 |
Description: | Case and dome temperatures for longwave fluxes are sporadically bad or noisy during these periods corresponding to additional noise in the longwave fluxes. Some checks can't be applied to the LW fluxes. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE16.s2:
|