DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120416.3 | SGP/QCRAD/E21 - Tracker out of alignment | sgpqcrad1longE21.s2, sgpqcrad1longE21.c2 | Incorrect |
D120510.10 | SGP/QCRAD/E25 - Noisy LWup Tc/Td | sgpqcrad1longE25.c2, sgpqcrad1longE25.s2 | Suspect |
D120510.16 | SGP/QCRAD/E3 - LWup dome temperature missing | sgpqcrad1longE3.c2, sgpqcrad1longE3.s2 | Suspect |
D120510.21 | SGP/QCRAD/E10 - Upwelling LW Tc/Td/flux questionable | sgpqcrad1longE10.s2 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/19/2008 | 2324 | 07/30/2008 | 1450 | Incorrect |
09/02/2005 | 2030 | 09/27/2005 | 2130 | Incorrect |
05/14/2001 | 1500 | 06/02/2001 | 0000 | Incorrect |
06/09/2008 | 0030 | 06/10/2008 | 1900 | Incorrect |
05/06/2000 | 0300 | 05/16/2000 | 0000 | Incorrect |
07/17/2003 | 1000 | 08/06/2003 | 1350 | Incorrect |
09/25/2008 | 1200 | 10/21/2008 | 2150 | Incorrect |
08/15/2008 | 1630 | 09/24/2008 | 1445 | Incorrect |
05/19/2003 | 1000 | 05/28/2003 | 1430 | Incorrect |
05/27/2008 | 1300 | 05/31/2008 | 1715 | Incorrect |
02/20/2004 | 1300 | 03/03/2004 | 1300 | Incorrect |
07/17/2000 | 0600 | 07/26/2000 | 2200 | Incorrect |
06/03/2003 | 1000 | 06/16/2003 | 1350 | Incorrect |
07/31/2008 | 1130 | 08/13/2008 | 1545 | Incorrect |
Subject: | SGP/QCRAD/E21 - Tracker out of alignment |
DataStreams: | sgpqcrad1longE21.s2, sgpqcrad1longE21.c2 |
Description: | Tracker appears to be out of alignment or inoperative during these times. The short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data appear to be bad and not recoverable. Some of these dates are documented in instrument DQRs: D030529.3, D030612.1, D030722.2, D040229.2, D080602.2, D080722.1 |
Measurements: | sgpqcrad1longE21.c2:
sgpqcrad1longE21.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/05/2000 | 1400 | 12/13/2000 | 1800 | Suspect |
Subject: | SGP/QCRAD/E25 - Noisy LWup Tc/Td |
DataStreams: | sgpqcrad1longE25.c2, sgpqcrad1longE25.s2 |
Description: | Case and dome temperatures for upwelling Longwave fluxes are noisy during this period corresponding to additional noise in the longwave fluxes. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE25.c2:
sgpqcrad1longE25.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/10/2000 | 0000 | 08/30/2000 | 2100 | Suspect |
Subject: | SGP/QCRAD/E3 - LWup dome temperature missing |
DataStreams: | sgpqcrad1longE3.c2, sgpqcrad1longE3.s2 |
Description: | Upwelling longwave dome temperatures are missing during this period, meaning not all checks/corrections can be made to LWup flux. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE3.s2:
sgpqcrad1longE3.c2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/01/2000 | 0000 | 01/01/2001 | 0000 | Suspect |
Subject: | SGP/QCRAD/E10 - Upwelling LW Tc/Td/flux questionable |
DataStreams: | sgpqcrad1longE10.s2 |
Description: | Case and dome temperatures for upwelling longwave flux are sporadically questionable during this period corresponding to additional noise in the longwave fluxes. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE10.s2:
|