DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D141215.2 | OLI/MFR/M1 - Shadow under tower impacting data | olimfr3mM1.a0 | Incorrect |
D150123.9 | OLI/AMC/M1 - Data missing from loss of power | oliamcM1.b1 | Missing |
D150526.2 | OLI/MFR/M1 - Data are missing | olimfr3mM1.a0 | Missing |
D150910.3 | OLI/RWP/M1 - Reprocess: Incorrect beam direction | oli915rwpwindmeanlowM1.a1 | Incorrect |
D160114.7 | OLI/GHG/M1 - Instrument setup probationary period | oliaosghgauxM1.a1 | Incorrect |
D160213.1 | OLI/GHG/M1 - Chiller failure | oliaosghgauxM1.a1 | Incorrect |
D160213.2 | OLI/GHG/M1 - Tower pump failure | oliaosghgauxM1.a1 | Suspect |
D160317.7 | OLI/GHG/M1 - Line pressure control failure | oliaosghgauxM1.a1 | Incorrect |
D160401.5 | OLI/DL/M1 - Instrument computer issue | olidlauxM1.a0 | Missing |
D160408.2 | OLI/GHG/M1 - Pressure controller failure | oliaosghgauxM1.a1 | Incorrect |
D160520.8 | OLI/DL/DLPROF/M1 - Reprocess: Incorrect serial number | olidlprofwind4newsM1.c1, olidlprofwstats4newsM1.c1 | Does not affect quality |
D160608.4 | OLI/DL/M1 - Instrument swap | olidlauxM1.a0 | Missing |
D160729.4 | OLI/RL/M1 - Data unavailable | olirllogM1.a0 | Missing |
D160811.2 | OLI/RL/M1 - Laser system problem | olirllogM1.a0 | Missing |
D160824.1 | OLI/GHG/M1 - Reprocess: oliaosghgauxM1.a1 | oliaosghgauxM1.a1 | Does not affect quality |
D160901.3 | OLI/CPC/M1 - Collection issue | oliaoscpcuM1.00 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/09/2014 | 2238 | 11/09/2014 | 2359 | Incorrect |
10/31/2014 | 2238 | 10/31/2014 | 2359 | Incorrect |
11/07/2014 | 2238 | 11/07/2014 | 2359 | Incorrect |
11/05/2014 | 2238 | 11/05/2014 | 2359 | Incorrect |
11/03/2014 | 2238 | 11/03/2014 | 2359 | Incorrect |
10/30/2014 | 2238 | 10/30/2014 | 2359 | Incorrect |
11/01/2014 | 2238 | 11/01/2014 | 2359 | Incorrect |
11/08/2014 | 2238 | 11/08/2014 | 2359 | Incorrect |
11/04/2014 | 2238 | 11/04/2014 | 2359 | Incorrect |
11/06/2014 | 2238 | 11/06/2014 | 2359 | Incorrect |
11/11/2014 | 2238 | 11/11/2014 | 2359 | Incorrect |
11/10/2014 | 2238 | 11/10/2014 | 2359 | Incorrect |
11/02/2014 | 2238 | 11/02/2014 | 2359 | Incorrect |
11/12/2014 | 2238 | 11/12/2014 | 2359 | Incorrect |
10/29/2014 | 2238 | 10/29/2014 | 2359 | Incorrect |
Subject: | OLI/MFR/M1 - Shadow under tower impacting data |
DataStreams: | olimfr3mM1.a0 |
Description: | We believe a shadow from a nearby structure, "the hangar", is affecting the MFR data. The suspected impact can generally be seen on clear days, e.g., 10/29, 11/02, 11/06, 11/09, and 11/12. Data on overcast or mostly cloudy days are likely OK. |
Measurements: | olimfr3mM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/06/2015 | 1430 | 01/08/2015 | 2000 | Missing |
Subject: | OLI/AMC/M1 - Data missing from loss of power |
DataStreams: | oliamcM1.b1 |
Description: | Data is missing due to logger battery voltage dropping below operational range of about 9 VDC, and the logger shutting down automatically in order to preserve battery life. Starting with 12/31/2014 04:00:00, AMF3 facility lost power due to failure of microturbines. The system was operating normally until the logger shut down at this time. |
Measurements: | oliamcM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/19/2015 | 1359 | 04/21/2015 | 1903 | Missing |
Subject: | OLI/MFR/M1 - Data are missing |
DataStreams: | olimfr3mM1.a0 |
Description: | Data are missing and cannot be recovered. |
Measurements: | olimfr3mM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/07/2015 | 0000 | 07/28/2015 | 0000 | Incorrect |
Subject: | OLI/RWP/M1 - Reprocess: Incorrect beam direction |
DataStreams: | oli915rwpwindmeanlowM1.a1 |
Description: | Incorrect beam direction results in incorrect data. Data will be reprocessed to correct for incorrect beam directions. |
Measurements: | oli915rwpwindmeanlowM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/31/2015 | 2256 | 06/28/2015 | 2359 | Incorrect |
Subject: | OLI/GHG/M1 - Instrument setup probationary period |
DataStreams: | oliaosghgauxM1.a1 |
Description: | Instrument setup and probationary period. Additional information: 2015-06-12 10:00:00,2015-06-22 05:59:59 - Multiport valve was 0; not powered on or connected to data acquisition software 2015-06-22 06:00:00,2015-06-28 02:59:59 - Site intermittent power loss |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/21/2015 | 0600 | 08/10/2015 | 1659 | Incorrect |
Subject: | OLI/GHG/M1 - Chiller failure |
DataStreams: | oliaosghgauxM1.a1 |
Description: | Chiller failure during this period |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/26/2015 | 1200 | 10/26/2015 | 1759 | Suspect |
Subject: | OLI/GHG/M1 - Tower pump failure |
DataStreams: | oliaosghgauxM1.a1 |
Description: | Tower pump failure |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/07/2016 | 0918 | 03/08/2016 | 0721 | Incorrect |
03/10/2016 | 0626 | 03/11/2016 | 2237 | Incorrect |
Subject: | OLI/GHG/M1 - Line pressure control failure |
DataStreams: | oliaosghgauxM1.a1 |
Description: | When the pressure controller is not controlling pressure adequately, co2/ch4 data cannot be used |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/22/2016 | 1633 | 03/25/2016 | 2137 | Missing |
Subject: | OLI/DL/M1 - Instrument computer issue |
DataStreams: | olidlauxM1.a0 |
Description: | The instrument computer did not boot up normally following the installation of security patches to the operating system. The problem was eventually resolved. |
Measurements: | olidlauxM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/25/2016 | 1700 | 04/08/2016 | 2100 | Incorrect |
Subject: | OLI/GHG/M1 - Pressure controller failure |
DataStreams: | oliaosghgauxM1.a1 |
Description: | Pressure controller failure causes CO2, CH4, and H2O data to be unusable |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/28/2014 | 0000 | 05/12/2016 | 1800 | Does not affect quality |
Subject: | OLI/DL/DLPROF/M1 - Reprocess: Incorrect serial number |
DataStreams: | olidlprofwind4newsM1.c1, olidlprofwstats4newsM1.c1 |
Description: | The instrument serial number that appears in the netcdf files global attributes is incorrect. The correct serial number is 0514-83. |
Measurements: | olidlprofwstats4newsM1.c1:
olidlprofwind4newsM1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/12/2016 | 1825 | 06/04/2016 | 1907 | Missing |
Subject: | OLI/DL/M1 - Instrument swap |
DataStreams: | olidlauxM1.a0 |
Description: | No data were acquired during this period because the instrument was in the process of being swapped out. Prior to 2016-05-12, the Oliktok Doppler lidar was a Halo Photonics StreamLine (S/N 0514-83). This was replaced with a Halo Photonics StreamLinePro (S/N 0116-109). |
Measurements: | olidlauxM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/31/2015 | 1201 | 07/14/2016 | 2324 | Missing |
Subject: | OLI/RL/M1 - Data unavailable |
DataStreams: | olirllogM1.a0 |
Description: | No data available |
Measurements: | olirllogM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/31/2016 | 1725 | 08/01/2016 | 2004 | Missing |
Subject: | OLI/RL/M1 - Laser system problem |
DataStreams: | olirllogM1.a0 |
Description: | The Raman lidar was down during the period in question due to issues with the laser system. |
Measurements: | olirllogM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/31/2015 | 0000 | 04/19/2016 | 2359 | Does not affect quality |
Subject: | OLI/GHG/M1 - Reprocess: oliaosghgauxM1.a1 |
DataStreams: | oliaosghgauxM1.a1 |
Description: | D160330.1 was recently completed. But that DQR did not specify rearchiving the oliaosghgauxM1.a1 data that is created when the oliaosghgM1.a1 data is created. We need to re-reprocess this date and this time store the oliaosghgauxM1.a1 data in the Archive. There will be some minor changes in the oliaosghgauxM1.a1 data. When NaN values are found in the raw data, those values will be replaced by values of -9999. |
Measurements: | oliaosghgauxM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/25/2016 | 1921 | 08/30/2016 | 1755 | Missing |
Subject: | OLI/CPC/M1 - Collection issue |
DataStreams: | oliaoscpcuM1.00 |
Description: | The LabVIEW VI collecting data from the CPC 3776 instrument stopped working and data were missed during this interval. |
Measurements: | oliaoscpcuM1.00:
|