DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160907.12 | ASI/CPC/M1 - Instrument testing | asiaoscpcfM1.b1 | Incorrect |
D160907.13 | ASI/CPC/M1 - Instrument maintence | asiaoscpcfM1.b1 | Incorrect |
D160907.14 | ASI/SMPS/M1 - Instrument maintence | asiaossmpsM1.a1 | Incorrect |
D161003.1 | ASI/SMPS/M1 - Small leak in the SMPS CPC | asiaossmpsM1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/03/2016 | 1314 | 09/03/2016 | 1327 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument testing |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During this time tests were being run with a HEPA filter on the inlet. Data is not representative of ambient aerosol concentrations |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1004 | 09/06/2016 | 1029 | Incorrect |
09/06/2016 | 1815 | 09/06/2016 | 1819 | Incorrect |
Subject: | ASI/CPC/M1 - Instrument maintence |
DataStreams: | asiaoscpcfM1.b1 |
Description: | During these intervals, connections were being removed from the CPC to prepare for it to take the place of the SMPS CPC. |
Measurements: | asiaoscpcfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/06/2016 | 1030 | 09/06/2016 | 1154 | Incorrect |
08/29/2016 | 1527 | 08/29/2016 | 1549 | Incorrect |
09/06/2016 | 1725 | 09/06/2016 | 1744 | Incorrect |
08/29/2016 | 1605 | 08/31/2016 | 1719 | Incorrect |
09/02/2016 | 1305 | 09/02/2016 | 1334 | Incorrect |
09/06/2016 | 1800 | 09/06/2016 | 1824 | Incorrect |
09/01/2016 | 1309 | 09/01/2016 | 1814 | Incorrect |
08/29/2016 | 1255 | 08/29/2016 | 1344 | Incorrect |
Subject: | ASI/SMPS/M1 - Instrument maintence |
DataStreams: | asiaossmpsM1.a1 |
Description: | During these time periods, a HEPA filter was connected to different portions of the system for investigation of leaks. |
Measurements: | asiaossmpsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/29/2016 | 1345 | 08/29/2016 | 1526 | Suspect |
08/29/2016 | 1550 | 08/29/2016 | 1604 | Suspect |
09/03/2016 | 1230 | 09/03/2016 | 1304 | Suspect |
05/19/2016 | 0000 | 08/29/2016 | 1254 | Suspect |
09/02/2016 | 1335 | 09/02/2016 | 1419 | Suspect |
Subject: | ASI/SMPS/M1 - Small leak in the SMPS CPC |
DataStreams: | asiaossmpsM1.a1 |
Description: | Periodic spikes were seen in data due to a small leak in the SMPS CPC. These spikes are identifiable and may be corrected in future data products. |
Measurements: | asiaossmpsM1.a1:
|