DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160531.1 | ASI/AOSMET/M1 - Reprocess: Wind sensor not aligned to north | asiaosmetM1.a1 | Incorrect |
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaosmetM1.a1 | Incorrect |
D161013.5 | ASI/AOSMET/M1/M1 - Computer disk failure | asiaosmetM1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/23/2016 | 0000 | 05/31/2016 | 1646 | Incorrect |
Subject: | ASI/AOSMET/M1 - Reprocess: Wind sensor not aligned to north |
DataStreams: | asiaosmetM1.a1 |
Description: | The wind sensor was not aligned to true north, resulting in wind direction data 16 degrees too high. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaosmetM1.a1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosmetM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2016 | 1500 | 10/13/2016 | 1800 | Does not affect quality |
Subject: | ASI/AOSMET/M1/M1 - Computer disk failure |
DataStreams: | asiaosmetM1.a1 |
Description: | Data are intermittently missing from 10/4 until the computer disk was replaced on 10/13. Data are completely missing 10/12 at 11:00 UTC until 10/13 at 18:00 UTC. |
Measurements: | asiaosmetM1.a1:
|