DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160129.4 | AWR/MWACR/M1 - Transmitter Down | awrmwacrM1.a1 | Incorrect |
D160407.5 | AWR/MWACR/M1 - Transmitter Down | awrmwacrM1.a1 | Incorrect |
D160407.6 | AWR/MWACR/M1 - Instrument Down | awrmwacrM1.a1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/28/2016 | 1658 | 01/28/2016 | 1806 | Incorrect |
Subject: | AWR/MWACR/M1 - Transmitter Down |
DataStreams: | awrmwacrM1.a1 |
Description: | The transmitter was down during this period |
Measurements: | awrmwacrM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/29/2016 | 1729 | 01/30/2016 | 0633 | Incorrect |
03/19/2016 | 0149 | 03/19/2016 | 2359 | Incorrect |
01/28/2016 | 1658 | 01/28/2016 | 1807 | Incorrect |
02/17/2016 | 1946 | 02/29/2016 | 1759 | Incorrect |
02/02/2016 | 1521 | 02/02/2016 | 2016 | Incorrect |
Subject: | AWR/MWACR/M1 - Transmitter Down |
DataStreams: | awrmwacrM1.a1 |
Description: | The MWACR was faulting during these periods causing the transmitter to turn off. Problem was a faulty HV modulator. |
Measurements: | awrmwacrM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/22/2016 | 1813 | 02/27/2016 | 2033 | Missing |
02/29/2016 | 1800 | 03/19/2016 | 0149 | Missing |
03/20/2016 | 0000 | 01/13/2017 | 2359 | Missing |
02/20/2016 | 0000 | 02/22/2016 | 0659 | Missing |
Subject: | AWR/MWACR/M1 - Instrument Down |
DataStreams: | awrmwacrM1.a1 |
Description: | The instrument was faulting and turning off at times. It was removed for repairs and will not be redeployed. |
Measurements: | awrmwacrM1.a1:
|