DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120522.10 | TWP/QCRAD/C1 - LWdn ventilator failed | twpqcrad1longC1.s2 | Suspect |
D140501.4 | TWP/QCRAD/C1 - Reprocessing Notice (RID 1241) | twpqcrad1longC1.s2 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/14/2008 | 0200 | 08/18/2008 | 0545 | Suspect |
Subject: | TWP/QCRAD/C1 - LWdn ventilator failed |
DataStreams: | twpqcrad1longC1.s2 |
Description: | LWdn flux may be questionable during this time. Described in instrument DQR D080826.3: During the time range specified the PIR2 output is questionable due to ventilator fan failure. Although the PIR is temperature compensated, data review indicates that the PIR2 didn't compare as well with PIR1. Typical differences during full sun are approximately 10 w/m2. During this time period differences of 10-14 W/m2 were noted. |
Measurements: | twpqcrad1longC1.s2:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/28/2011 | 0000 | 10/05/2013 | 2359 | Does not affect quality |
Subject: | TWP/QCRAD/C1 - Reprocessing Notice (RID 1241) |
DataStreams: | twpqcrad1longC1.s2 |
Description: | See below for details of reprocessing that was completed on 20141217. Original DQR category: 0 (Black, Missing). ------------------------------------------- Upwelling data is missing |
Measurements: | twpqcrad1longC1.s2:
|