DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140326.2 | TMP/NEPHELOMETER/S1 - Reprocess: Incorrect values for impactor_setting | tmpaosnephwetS1.b1 | Incorrect |
D140826.4 | TMP/NEPHELOMETER/S1 - Nephelometers were connected in series | tmpaosnephwetS1.b1 | Suspect |
D140924.2 | TMP/NEPHELOMETER/S1 - High humidity effects | tmpaosnephwetS1.b1 | Incorrect |
D150701.20 | TMP/NEPHELOMETER/S1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 | tmpaosnephwetS1.b1 | Does not affect quality |
D151112.3 | TMP/AETH/NEPHELOMETER/PASS/PSAP/S1 - Impactor state occasionally incorrect | tmpaosnephwetS1.b1 | Does not affect quality |
D151112.4 | TMP/AETH/NEPHELOMETER/PASS/PSAP/S1 - Mislabeled particle sizes | tmpaosnephwetS1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/30/2014 | 1500 | 03/07/2014 | 1558 | Incorrect |
Subject: | TMP/NEPHELOMETER/S1 - Reprocess: Incorrect values for impactor_setting |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | Particle sizes incorrect with impactor particle size cut settings 10 and 1 reversed. |
Measurements: | tmpaosnephwetS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/05/2014 | 0700 | 06/26/2014 | 1320 | Suspect |
Subject: | TMP/NEPHELOMETER/S1 - Nephelometers were connected in series |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | Nephelometers were connected in series and run as ambient nephelometers due to an O-ring failure. |
Measurements: | tmpaosnephwetS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/26/2014 | 2000 | 09/15/2014 | 0459 | Incorrect |
Subject: | TMP/NEPHELOMETER/S1 - High humidity effects |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | The present humidigraph software cannot control the humidity and in high humidity environments it can exceed 95% resulting in water in the wet nephelometer. |
Measurements: | tmpaosnephwetS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/30/2014 | 0000 | 09/15/2014 | 2359 | Does not affect quality |
Subject: | TMP/NEPHELOMETER/S1 - Reprocess: reprocessing from raw data in preparation for |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | End to end reprocessing to generate new a1 datastreams is needed so that when the harmonization software is released to production, we will be ready with the a1 datastreams that are input to the subsequent processing software. |
Measurements: | tmpaosnephwetS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2014 | 0544 | 06/22/2014 | 0612 | Does not affect quality |
Subject: | TMP/AETH/NEPHELOMETER/PASS/PSAP/S1 - Impactor state occasionally incorrect |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | The computer controlling the impactor state was rebooting every 15 minutes causing periodic gaps in the data and interrupting the normal sequence of states. |
Measurements: | tmpaosnephwetS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/13/2013 | 2000 | 03/07/2014 | 1558 | Does not affect quality |
Subject: | TMP/AETH/NEPHELOMETER/PASS/PSAP/S1 - Mislabeled particle sizes |
DataStreams: | tmpaosnephwetS1.b1 |
Description: | Particle sizes as indicated with impactor are incorrect with particle size cuts 10 and 1 reversed. |
Measurements: | tmpaosnephwetS1.b1:
|