DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140527.1 | MAO/CSPHOT/M1 - Obstruction | maocsphotaotfiltM1.a1 | Suspect |
D141008.1 | MAO/CSPHOT/M1 - Instrument and power failure | maocsphotaotfiltM1.a1 | Missing |
D141031.5 | MAO/CSPHOT/M1 - 340 nm Filter Problem | maocsphotaotfiltM1.a1 | Suspect |
D150331.1 | MAO/CSPHOT/M1 - Instrument obstruction | maocsphotaotfiltM1.a1 | Suspect |
D151002.7 | MAO/CSPHOT/M1 - Instrument Problem - Obstruction | maocsphotaotfiltM1.a1 | Suspect |
D160608.8 | MAO/CSPHOT/M1 - No post-deployment calibration | maocsphotaotfiltM1.a1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/08/2014 | 0000 | 04/30/2014 | 2359 | Suspect |
Subject: | MAO/CSPHOT/M1 - Obstruction |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | An obstruction in the collimator caused a discrepancy in the 1640 channel. |
Measurements: | maocsphotaotfiltM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/20/2014 | 0000 | 08/25/2014 | 2359 | Missing |
07/18/2014 | 0000 | 08/18/2014 | 2359 | Missing |
09/19/2014 | 0000 | 10/04/2014 | 2359 | Missing |
06/29/2014 | 0000 | 07/13/2014 | 2359 | Missing |
Subject: | MAO/CSPHOT/M1 - Instrument and power failure |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | Data are missing and are unrecoverable. |
Measurements: | maocsphotaotfiltM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/15/2014 | 0000 | 12/03/2014 | 2359 | Suspect |
Subject: | MAO/CSPHOT/M1 - 340 nm Filter Problem |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | The 340 nm channel is suspect due to a filter issue. |
Measurements: | maocsphotaotfiltM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/20/2015 | 0000 | 03/25/2015 | 2359 | Suspect |
Subject: | MAO/CSPHOT/M1 - Instrument obstruction |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | An obstruction in the collimator caused a discrepancy in the 1640 channel. |
Measurements: | maocsphotaotfiltM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/22/2015 | 0000 | 09/02/2015 | 2359 | Suspect |
Subject: | MAO/CSPHOT/M1 - Instrument Problem - Obstruction |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | Data are questionable due to obstruction in collimator tubes. |
Measurements: | maocsphotaotfiltM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/12/2014 | 1710 | 01/28/2015 | 1710 | Suspect |
Subject: | MAO/CSPHOT/M1 - No post-deployment calibration |
DataStreams: | maocsphotaotfiltM1.a1 |
Description: | The instrument deployed during this period was lost on the way back to the US. Therefore, it could not be recalibrate as is customary. The differences between pre- and post-deployment calibrations for such a short deployment are typically quite small, 1%. But occasionally there are dramatic shifts that we now won't know about. |
Measurements: | maocsphotaotfiltM1.a1:
|