DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160621.4 | ASI/AOS/M1 - Manifold blower not working | asiaosimpactorM1.b1 | Incorrect |
D160708.6 | ASI/AOS/M1 - Incorrect readout of Impactor state in 10-um setting not working | asiaosimpactorM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2016 | 0042 | 06/16/2016 | 1345 | Incorrect |
Subject: | ASI/AOS/M1 - Manifold blower not working |
DataStreams: | asiaosimpactorM1.b1 |
Description: | Manifold blower not working, instrument sample flows either zero or different than specified. Sampling through the stack not isokinetic. |
Measurements: | asiaosimpactorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/21/2016 | 1100 | 07/20/2016 | 1807 | Incorrect |
Subject: | ASI/AOS/M1 - Incorrect readout of Impactor state in 10-um setting not working |
DataStreams: | asiaosimpactorM1.b1 |
Description: | The limit switch in the 10-um position was not working. Operator found a loose wire and resoldered. |
Measurements: | asiaosimpactorM1.b1:
|