DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140530.2 | MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet | maoaossmpsS1.a1 | Suspect |
D151203.4 | MAO/SMPS/S1 - Reprocess: SMPS data | maoaossmpsS1.a1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/07/2014 | 1430 | 05/13/2014 | 2043 | Suspect |
Subject: | MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet |
DataStreams: | maoaossmpsS1.a1 |
Description: | The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the SMPS, which is nominally 1 lpm. This decrease in the SMPS sample flow rate results in an underestimation of the ambient size distribution and integrated number concentration. |
Measurements: | maoaossmpsS1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/29/2014 | 0000 | 12/28/2015 | 0000 | Does not affect quality |
Subject: | MAO/SMPS/S1 - Reprocess: SMPS data |
DataStreams: | maoaossmpsS1.a1 |
Description: | No affect on data quality |
Measurements: | maoaossmpsS1.a1:
|