DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120420.8 | SGP/QCRAD/C1 - NIP failure | sgpqcrad1longC1.s2 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2011 | 0000 | 06/24/2011 | 1740 | Incorrect |
Subject: | SGP/QCRAD/C1 - NIP failure |
DataStreams: | sgpqcrad1longC1.s2 |
Description: | The short_direct_normal (NIP instrument) cable failed during this time period. Data are not recoverable. Users can derive an estimate of direct normal values using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) Documented in instrument DQR D110628.2 |
Measurements: | sgpqcrad1longC1.s2:
|