DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D090213.3 | HFE/MWRHF/M1 - Calibration updates | hfemwrhfM1.b1 | Does not affect quality |
D090213.4 | HFE/MWRHF/M1 - Rain sensor failure | hfemwrhfM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/09/2008 | 0500 | 12/28/2008 | 2359 | Does not affect quality |
Subject: | HFE/MWRHF/M1 - Calibration updates |
DataStreams: | hfemwrhfM1.b1 |
Description: | The MWRHF calibration algorithm updates the calibration parameters every time a new successful tip curve is acquired. This results in small jumps (1-2 K) in the brightness temperatures each time calibration is updated. During times of frequent calibrations the data appear discontinuous with short data segments interrupted by small jumps. |
Measurements: | hfemwrhfM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/28/2008 | 0000 | 08/04/2008 | 0300 | Incorrect |
Subject: | HFE/MWRHF/M1 - Rain sensor failure |
DataStreams: | hfemwrhfM1.b1 |
Description: | The rain sensor failed and the rain flag was always on. When the rain flag is on the radiometer does not perform tip curves. The rain sensor was replaced 8/4. |
Measurements: | hfemwrhfM1.b1:
|