DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160722.3 | ASI/MPL/M1 - instrument | asimplpolfsM1.b1 | Missing |
D160914.1 | ASI/CEIL/M1 - data communication | asiceilM1.b1 | Suspect |
D161019.2 | ASI/MPL/M1 - Reprocess: New overlap correction | asimplpolfsM1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/01/2016 | 0000 | 07/02/2016 | 1305 | Missing |
Subject: | ASI/MPL/M1 - instrument |
DataStreams: | asimplpolfsM1.b1 |
Description: | Later than planned deployment of the instrument. |
Measurements: | asimplpolfsM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/12/2016 | 2000 | 08/14/2016 | 0459 | Suspect |
Subject: | ASI/CEIL/M1 - data communication |
DataStreams: | asiceilM1.b1 |
Description: | The data were intermittent, especially on 7/19/2016, due to poor serial communication. The cable was repaired on 7/21 and the serial device server and instrument were power-cycled on 7/28. |
Measurements: | asiceilM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/02/2016 | 1300 | 10/27/2016 | 2359 | Does not affect quality |
Subject: | ASI/MPL/M1 - Reprocess: New overlap correction |
DataStreams: | asimplpolfsM1.b1 |
Description: | Vendor has recently provided an overlap correction which needs to replace the overlap currently being included in the data files. This does not affect MPL data quality. |
Measurements: | asimplpolfsM1.b1:
|