DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D150123.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D150309.4 | ENA/GNDRAD/C1 - Instrument problem | enagndrad60sC1.b1 | Suspect |
D150327.7 | OLI/GNDRAD/M1 - Environmental | oligndrad60sM1.b1 | Incorrect |
D151102.3 | OLI/GNDRAD/M1 - PIR cable connector failure | oligndrad60sM1.b1 | Incorrect |
D160427.7 | ENA/C1 - Reprocess: Slightly incorrect location fields | enagndrad60sC1.b1 | Suspect |
D160610.14 | MCQ/GNDRAD/S1 - IRT not installed | mcqgndrad60sS1.b1 | Incorrect |
D160624.2 | NSA/GNDRAD/C1 - IRT wiring | nsagndrad60sC1.b1 | Incorrect |
D160927.1 | MCQ/GNDRAD/S1 - Instrument | mcqgndrad60sS1.b1 | Incorrect |
D170110.1 | OLI/M1 - Severe Storm | oligndrad60sM1.b1 | Does not affect quality |
D170116.4 | OLI/GNDRAD/M1 - power outage | oligndrad60sM1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/12/2014 | 2230 | 01/15/2015 | 0030 | Incorrect |
12/31/2014 | 0300 | 01/09/2015 | 0500 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | GNDRAD system not operating correctly during affected time periods due to power outage. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/08/2014 | 0000 | 02/21/2015 | 1425 | Suspect |
Subject: | ENA/GNDRAD/C1 - Instrument problem |
DataStreams: | enagndrad60sC1.b1 |
Description: | Intermittent noise in up_long_hemisp plots. |
Measurements: | enagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2015 | 1916 | 03/08/2015 | 0043 | Incorrect |
Subject: | OLI/GNDRAD/M1 - Environmental |
DataStreams: | oligndrad60sM1.b1 |
Description: | AC power out during affected time range affecting ventilator fans and heaters for up_short_hemisp, and up_long_hemisp radiometers. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/22/2015 | 0000 | 10/25/2015 | 0321 | Incorrect |
Subject: | OLI/GNDRAD/M1 - PIR cable connector failure |
DataStreams: | oligndrad60sM1.b1 |
Description: | During the specified time period upwelling PIR cable/connector problems were being corrected. |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/29/2013 | 0941 | 04/30/2016 | 2359 | Suspect |
Subject: | ENA/C1 - Reprocess: Slightly incorrect location fields |
DataStreams: | enagndrad60sC1.b1 |
Description: | Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be corrected. The respective GRW and ENA lat, lon, and alt values are as follows: GRW lat = 39.0911 ; ENA lat = 39.0916 ; GRW lon = -28.0297 ; ENA lon = -28.0257 ; GRW alt = 26 ; ENA alt = 30.48 ; |
Measurements: | enagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/03/2016 | 0543 | 03/31/2018 | 2359 | Incorrect |
Subject: | MCQ/GNDRAD/S1 - IRT not installed |
DataStreams: | mcqgndrad60sS1.b1 |
Description: | An IRT was not installed so the datalogger sampled a floating signal. |
Measurements: | mcqgndrad60sS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/24/2016 | 1835 | 06/03/2016 | 1910 | Incorrect |
Subject: | NSA/GNDRAD/C1 - IRT wiring |
DataStreams: | nsagndrad60sC1.b1 |
Description: | When maintenance was performed on the GNDRAD tower, the IRT signal cable was re-connected incorrectly. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/03/2016 | 0600 | 08/15/2016 | 0630 | Incorrect |
Subject: | MCQ/GNDRAD/S1 - Instrument |
DataStreams: | mcqgndrad60sS1.b1 |
Description: | Excessive noise in GNDRAD upwelling longwave during affected time period. |
Measurements: | mcqgndrad60sS1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/05/2017 | 1000 | 01/08/2017 | 1200 | Does not affect quality |
Subject: | OLI/M1 - Severe Storm |
DataStreams: | oligndrad60sM1.b1 |
Description: | The site lost power during a severe winter storm. Operators on site were not able to get out and restart the generators to bring things back up until 1/7/2017. Most of the instruments were brought up at this time, but some problems remained until 1/8/2017 |
Measurements: | oligndrad60sM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/05/2017 | 1815 | 01/07/2017 | 1630 | Missing |
Subject: | OLI/GNDRAD/M1 - power outage |
DataStreams: | oligndrad60sM1.b1 |
Description: | Blizzard conditions resulted in a power outage. |
Measurements: | oligndrad60sM1.b1:
|