DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160919.1 | ASI/SASZE/M1 - Spectrometer chiller temperature variations | asisaszefilterbandsM1.a1 | Suspect |
D161209.4 | ASI/SASZE/M1 - Software hang, no data | asisaszefilterbandsM1.a1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/03/2016 | 1655 | 07/21/2016 | 2359 | Suspect |
Subject: | ASI/SASZE/M1 - Spectrometer chiller temperature variations |
DataStreams: | asisaszefilterbandsM1.a1 |
Description: | Spectrometer chiller temperature variations caused by an incorrectly placed sensor are apparent as correlated modulations in the signal from the UV-VIS CCD spectrometer. The magnitude of signal oscillation is small and is only apparent under clear sky for wavelengths < 450 nm. The variations were reduced to negligible levels by repositioning the sensor on Aug. 21, 2016. |
Measurements: | asisaszefilterbandsM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/17/2016 | 1410 | 11/28/2016 | 1859 | Missing |
10/08/2016 | 2303 | 10/10/2016 | 1744 | Missing |
08/31/2016 | 1300 | 09/01/2016 | 1609 | Missing |
09/05/2016 | 1309 | 10/07/2016 | 1702 | Missing |
09/03/2016 | 1301 | 09/05/2016 | 1241 | Missing |
Subject: | ASI/SASZE/M1 - Software hang, no data |
DataStreams: | asisaszefilterbandsM1.a1 |
Description: | No data is generated while the software hung. |
Measurements: | asisaszefilterbandsM1.a1:
|