DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D151030.1 | ENA/ACSM/C1 - Instrument waiting for spare parts | enaaosacsmC1.b1 | Incorrect |
D160421.1 | ENA/ACSM/C1 - Missing Data | enaaosacsmC1.b1 | Missing |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/14/2015 | 2006 | 02/19/2016 | 1501 | Incorrect |
Subject: | ENA/ACSM/C1 - Instrument waiting for spare parts |
DataStreams: | enaaosacsmC1.b1 |
Description: | The ACSM was out of service while waiting for a replacement valve and a replacement turbo molecular vacuum pump. The data acquisition was stopped and the instrument put in standby mode. |
Measurements: | enaaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/18/2013 | 0000 | 07/10/2014 | 1010 | Missing |
Subject: | ENA/ACSM/C1 - Missing Data |
DataStreams: | enaaosacsmC1.b1 |
Description: | Data is missing due to a failed filter switching valve. The replacement valve was delivered in June 2014. There were tuning issues with the instrument and the instrument was successfully tuned on July 10, 2014, and started generating good data. |
Measurements: | enaaosacsmC1.b1:
|