Data Quality Reports for Session: 183757 User: yann1blanc Completed: 02/24/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D071203.3SGP/MPL/C1 - Scope temperature incorrectsgpmplpolavgC1.s1, sgpmplpolavgC1.c1, sgpmplpolC1.b1Incorrect


DQRID : D071203.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/29/2007190008/06/20071900Incorrect
Subject:
SGP/MPL/C1 - Scope temperature incorrect
DataStreams:sgpmplpolavgC1.s1, sgpmplpolavgC1.c1, sgpmplpolC1.b1
Description:
The scope_temp started to jump around on 6/29 at about 19:00 GMT between 24-32C.  Then 
~03:00 GMT on 6/30 the temperature exceeded the MAX flag value of 40C and was fluctuating 
between 150-225C.  

Temp probe connection was found to be loose and was reseated 20070806.

Ignore scope temp data and assume values are valid.
Measurements:sgpmplpolavgC1.s1:
  • scope_temp

sgpmplpolavgC1.c1:
  • scope_temp

sgpmplpolC1.b1:
  • scope_temp


Back To Table of Contents