Data Quality Reports for Session: 183824 User: pmarin Completed: 02/28/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D130705.2SGP/AOS/C1 - Missing datasgpnoaaaosccn100C1.b1Missing
D140205.4SGP/CCN/C1 - CCN shutdown for repairsgpaosccnavgC1.c1Missing
D140929.2SGP/AOS/C1 - Frequent power and communication loss problemssgpnoaaaosccn100C1.b1Does not affect quality
D140929.3SGP/CCN/C1 - Frequent power and communication loss problemssgpaosccnavgC1.c1Does not affect quality
D140929.5SGP/CCN/C1 - CCN presure sensor failuresgpaosccnavgC1.c1Incorrect
D141017.5SGP/CCN/C1 - Incorrect supersaturation stepping methodsgpaosccnavgC1.c1Incorrect
D141031.7SGP/AOS/C1 - New UPS installedsgpnoaaaosccn100C1.b1Missing
D141208.1SGP/CCN/C1 - CCN signal is low from possible leak and decline in laser powersgpaosccnavgC1.c1Suspect
D141211.1SGP/CCN/C1 - Incorrect temperature gains or settingssgpaosccnavgC1.c1Incorrect
D150306.2SGP/CCN/C1 - Instrument under counting at lowest super saturation valuesgpaosccnavgC1.c1Suspect
D150413.1SGP/CCN/C1 - Removed for calibrationsgpaosccnavgC1.c1Missing
D151103.3SGP/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/C1 - System MaintenancesgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1Incorrect
D151210.2SGP/AOS/C1 - Pump and blower failuresgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1Incorrect
D151210.5SGP/AOS/C1 - Power outagesgpnoaaaosccn100C1.b1Incorrect
D160104.7SGP Site Power Outage Due to Winter Stormsgpnoaaaosccn100C1.b1, sgpaosccnavgC1.c1Does not affect quality
D160406.2SGP/AOS/C1 - Wildfire smoke plumes advecting over facilitysgpnoaaaosccn100C1.b1Does not affect quality


DQRID : D130705.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/13/2013180006/17/20131506Missing
Subject:
SGP/AOS/C1 - Missing data
DataStreams:sgpnoaaaosccn100C1.b1
Description:
Data are missing and unrecoverable.
Measurements:sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • CCN_laser_current
  • Mean number concentration of N_CCN(N_CCN)
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_temperature_std
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_t_read_TEC3
  • CCN_temp_unstable_flag
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • time
  • CCN_T_OPC
  • Altitude above mean sea level(alt)
  • CCN_temperature_gradient
  • time_offset
  • East longitude(lon)
  • base_time
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • North latitude(lat)
  • CCN_T_sample


Back To Table of Contents

DQRID : D140205.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/04/2013153010/19/20130100Missing
Subject:
SGP/CCN/C1 - CCN shutdown for repair
DataStreams:sgpaosccnavgC1.c1
Description:
A thermal electric cooler (TEC) was found to be not working. The instrument was shutdown 
until the TEC could be replaced.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • N_CPC_std
  • North latitude(lat)
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • time_offset
  • Mean number concentration of N_CCN(N_CCN)
  • East longitude(lon)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • Altitude above mean sea level(alt)
  • time
  • base_time


Back To Table of Contents

DQRID : D140929.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/12/2014061407/21/20141340Does not affect quality
Subject:
SGP/AOS/C1 - Frequent power and communication loss problems
DataStreams:sgpnoaaaosccn100C1.b1
Description:
There were frequent power and communication losses over the time period. The problem was 
resolved after power was redistributed and all data acquisition components were replaced.
Measurements:sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • CCN_laser_current
  • Mean number concentration of N_CCN(N_CCN)
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_temperature_std
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_t_read_TEC3
  • CCN_temp_unstable_flag
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • time
  • CCN_T_OPC
  • Altitude above mean sea level(alt)
  • CCN_temperature_gradient
  • time_offset
  • East longitude(lon)
  • base_time
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • North latitude(lat)
  • CCN_T_sample


Back To Table of Contents

DQRID : D140929.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/12/2014061407/15/20142250Does not affect quality
Subject:
SGP/CCN/C1 - Frequent power and communication loss problems
DataStreams:sgpaosccnavgC1.c1
Description:
There were frequent power and communication losses over the time period. The problem was 
resolved after power was redistributed and all data acquisition components were replaced.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • N_CPC_std
  • North latitude(lat)
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • time_offset
  • Mean number concentration of N_CCN(N_CCN)
  • East longitude(lon)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • Altitude above mean sea level(alt)
  • time
  • base_time


Back To Table of Contents

DQRID : D140929.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/15/2014225008/12/20141400Incorrect
Subject:
SGP/CCN/C1 - CCN presure sensor failure
DataStreams:sgpaosccnavgC1.c1
Description:
Sheath flow reading declined and couldn't be adjusted. Unit was shipped back to mentor for 
repair and calibration.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • N_CPC_std
  • supersaturation_calculated
  • Mean number concentration of N_CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • column_temperature_gradient_measured
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_setting
  • droplet_size


Back To Table of Contents

DQRID : D141017.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/30/2012000006/11/20132359Incorrect
Subject:
SGP/CCN/C1 - Incorrect supersaturation stepping method
DataStreams:sgpaosccnavgC1.c1
Description:
The supersaturation stepping method over this period is not consistent with the ccnavg VAP 
expectations. Therefore, the averaging of the fields done by the CCNAVG VAP are 
incorrect - the output of the CCNAVG VAP for all fields is bad over this entire period.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • N_CPC_std
  • supersaturation_calculated
  • Mean number concentration of N_CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • column_temperature_gradient_measured
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_setting
  • droplet_size


Back To Table of Contents

DQRID : D141031.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/23/2014195410/23/20142125Missing
Subject:
SGP/AOS/C1 - New UPS installed
DataStreams:sgpnoaaaosccn100C1.b1
Description:
System down while new UPS is installed
Measurements:sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • CCN_laser_current
  • Mean number concentration of N_CCN(N_CCN)
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_temperature_std
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_t_read_TEC3
  • CCN_temp_unstable_flag
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • time
  • CCN_T_OPC
  • Altitude above mean sea level(alt)
  • CCN_temperature_gradient
  • time_offset
  • East longitude(lon)
  • base_time
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • North latitude(lat)
  • CCN_T_sample


Back To Table of Contents

DQRID : D141208.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/15/2009160008/03/20090000Suspect
Subject:
SGP/CCN/C1 - CCN signal is low from possible leak and decline in laser power
DataStreams:sgpaosccnavgC1.c1
Description:
CCN was returned with new laser. Signal was unusually low and flow calibration had changed 
substantially. A leak is suspect as well as declining laser power as new laser was 
defective.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • N_CPC_std
  • supersaturation_calculated
  • Mean number concentration of N_CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • column_temperature_gradient_measured
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_setting
  • droplet_size


Back To Table of Contents

DQRID : D141211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/22/2007220010/23/20071515Incorrect
Subject:
SGP/CCN/C1 - Incorrect temperature gains or settings
DataStreams:sgpaosccnavgC1.c1
Description:
New software was installed with instrument and temperature gains didn't transfer 
correctly. Super saturation values are incorrect and instrument tended to overheat in an attempt 
to reach set point values.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • N_CPC_std
  • supersaturation_calculated
  • Mean number concentration of N_CCN(N_CCN)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • column_temperature_gradient_measured
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_setting
  • droplet_size


Back To Table of Contents

DQRID : D150306.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/11/2013192604/08/20151730Suspect
Subject:
SGP/CCN/C1 - Instrument under counting at lowest super saturation value
DataStreams:sgpaosccnavgC1.c1
Description:
The slow temperature decay of the OPC induced droplet evaporation and caused an under 
counting of the droplets at the lowest %ss, 0.15.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_setpoint
  • supersaturation_calculated


Back To Table of Contents

DQRID : D150413.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/11/2014070003/15/20152045Missing
Subject:
SGP/CCN/C1 - Removed for calibration
DataStreams:sgpaosccnavgC1.c1
Description:
Instrument was removed for calibration and scanning evaluation.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • N_CPC_std
  • North latitude(lat)
  • column_temperature_gradient_measured_std
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • supersaturation_setpoint
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • time_offset
  • Mean number concentration of N_CCN(N_CCN)
  • East longitude(lon)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • Altitude above mean sea level(alt)
  • time
  • base_time


Back To Table of Contents

DQRID : D151103.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/20/2015204410/24/20152300Incorrect
Subject:
SGP/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/C1 - System Maintenance
DataStreams:sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1
Description:
The system was offline for cleaning and calibration.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • Altitude above mean sea level(alt)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • time
  • base_time
  • N_CPC_std
  • North latitude(lat)
  • column_temperature_gradient_measured_std
  • supersaturation_setpoint
  • time_offset
  • Mean number concentration of N_CCN(N_CCN)
  • East longitude(lon)

sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • CCN_laser_current
  • Mean number concentration of N_CCN(N_CCN)
  • CCN_temperature_std
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_t_read_TEC3
  • CCN_temp_unstable_flag
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • time
  • CCN_T_OPC
  • Altitude above mean sea level(alt)
  • time_offset
  • CCN_temperature_gradient
  • base_time
  • East longitude(lon)
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • North latitude(lat)
  • CCN_T_sample


Back To Table of Contents

DQRID : D151210.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/07/2015155212/08/20151649Incorrect
Subject:
SGP/AOS/C1 - Pump and blower failure
DataStreams:sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1
Description:
The fuse on the power strip with the pump and blower blew. The blower motor seized and the 
blower had to be replaced.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • N_CPC_std
  • column_temperature_gradient_measured_std
  • supersaturation_setpoint
  • Mean number concentration of N_CCN(N_CCN)

sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • CCN_laser_current
  • Mean number concentration of N_CCN(N_CCN)
  • CCN_temperature_std
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_temp_unstable_flag
  • CCN_t_read_TEC3
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • CCN_T_OPC
  • CCN_temperature_gradient
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • CCN_T_sample


Back To Table of Contents

DQRID : D151210.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/17/2015042511/17/20151600Incorrect
Subject:
SGP/AOS/C1 - Power outage
DataStreams:sgpnoaaaosccn100C1.b1
Description:
Power outage, no data collected.
Measurements:sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • Mean number concentration of N_CCN(N_CCN)
  • CCN_laser_current
  • CCN_temperature_std
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_super_saturation_calculated
  • CCN_Q_sheath
  • CCN_T_nafion
  • CCN_proportional_valve_voltage
  • CCN_t_read_TEC1
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_t_read_TEC2
  • CCN_temp_unstable_flag
  • CCN_t_read_TEC3
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_Q_sample
  • CCN_P_sample
  • CCN_T_sample
  • CCN_T_OPC
  • CCN_temperature_gradient


Back To Table of Contents

DQRID : D160104.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/26/2015000001/01/20162359Does not affect quality
Subject:
SGP Site Power Outage Due to Winter Storm
DataStreams:sgpnoaaaosccn100C1.b1, sgpaosccnavgC1.c1
Description:
On Saturday and Sunday, December 26 and 27th a powerful winter storm (Goliath) brought 
sleet and freezing rain to much of Oklahoma with wind gusts of 45 MPH at times. The result 
was very very slick roads and mass power outages.
 
The SGP central facility lost power at 8:55 PM CDT on Sunday, December 27th.  
The power was restored on 7:56 PM CDT on Wednesday, December 30th
 
Site Emergency procedures were activated and Technicians drained water out of instrument 
reservoirs and other measures as needed, early Monday morning.
 
On Wednesdays as roads and power started was restored to much of the areas east of I-35 , 
Site Technicians performed maintenance on several Extended Facilities.

During this period, data are potentially missing, incorrect, or suspect.  Carefully 
inspect the data before use.
Measurements:sgpaosccnavgC1.c1:
  • supersaturation_calculated
  • column_temperature_gradient_measured
  • column_temperature_gradient_setting
  • droplet_size
  • N_CCN_std
  • Condensation particle concentration number(N_CPC)
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • Altitude above mean sea level(alt)
  • time
  • base_time
  • N_CPC_std
  • North latitude(lat)
  • column_temperature_gradient_measured_std
  • supersaturation_setpoint
  • time_offset
  • Mean number concentration of N_CCN(N_CCN)
  • East longitude(lon)

sgpnoaaaosccn100C1.b1:
  • CCN_ss_set
  • Mean number concentration of N_CCN(N_CCN)
  • CCN_laser_current
  • CCN_temperature_std
  • AOS bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)
  • CCN_T_nafion
  • CCN_Q_sheath
  • CCN_t_read_TEC1
  • CCN_t_read_TEC2
  • CCN_t_read_TEC3
  • CCN_temp_unstable_flag
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • CCN_P_sample
  • time
  • CCN_T_OPC
  • Altitude above mean sea level(alt)
  • time_offset
  • CCN_temperature_gradient
  • East longitude(lon)
  • base_time
  • CCN_super_saturation_calculated
  • CCN_proportional_valve_voltage
  • CCN_T_inlet
  • CCN_first_stage_monitor_voltage
  • CCN_Q_sample
  • North latitude(lat)
  • CCN_T_sample


Back To Table of Contents

DQRID : D160406.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/18/2016040003/18/20161100Does not affect quality
03/25/2016010003/25/20160700Does not affect quality
Subject:
SGP/AOS/C1 - Wildfire smoke plumes advecting over facility
DataStreams:sgpnoaaaosccn100C1.b1
Description:
Smoke from nearby wildfires advected through SGP C1 during the listed affected time 
ranges, resulting in high signals observable in many of the listed AOS datastreams.
Measurements:sgpnoaaaosccn100C1.b1:
  • AOS CCN droplet count by bin size(N_CCN_dN)
  • Mean number concentration of N_CCN(N_CCN)


Back To Table of Contents



END OF DATA