DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D130705.2 | SGP/AOS/C1 - Missing data | sgpnoaaaosccn100C1.b1 | Missing |
D140205.4 | SGP/CCN/C1 - CCN shutdown for repair | sgpaosccnavgC1.c1 | Missing |
D140929.2 | SGP/AOS/C1 - Frequent power and communication loss problems | sgpnoaaaosccn100C1.b1 | Does not affect quality |
D140929.3 | SGP/CCN/C1 - Frequent power and communication loss problems | sgpaosccnavgC1.c1 | Does not affect quality |
D140929.5 | SGP/CCN/C1 - CCN presure sensor failure | sgpaosccnavgC1.c1 | Incorrect |
D141017.5 | SGP/CCN/C1 - Incorrect supersaturation stepping method | sgpaosccnavgC1.c1 | Incorrect |
D141031.7 | SGP/AOS/C1 - New UPS installed | sgpnoaaaosccn100C1.b1 | Missing |
D141208.1 | SGP/CCN/C1 - CCN signal is low from possible leak and decline in laser power | sgpaosccnavgC1.c1 | Suspect |
D141211.1 | SGP/CCN/C1 - Incorrect temperature gains or settings | sgpaosccnavgC1.c1 | Incorrect |
D150306.2 | SGP/CCN/C1 - Instrument under counting at lowest super saturation value | sgpaosccnavgC1.c1 | Suspect |
D150413.1 | SGP/CCN/C1 - Removed for calibration | sgpaosccnavgC1.c1 | Missing |
D151103.3 | SGP/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/C1 - System Maintenance | sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1 | Incorrect |
D151210.2 | SGP/AOS/C1 - Pump and blower failure | sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1 | Incorrect |
D151210.5 | SGP/AOS/C1 - Power outage | sgpnoaaaosccn100C1.b1 | Incorrect |
D160104.7 | SGP Site Power Outage Due to Winter Storm | sgpnoaaaosccn100C1.b1, sgpaosccnavgC1.c1 | Does not affect quality |
D160406.2 | SGP/AOS/C1 - Wildfire smoke plumes advecting over facility | sgpnoaaaosccn100C1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/13/2013 | 1800 | 06/17/2013 | 1506 | Missing |
Subject: | SGP/AOS/C1 - Missing data |
DataStreams: | sgpnoaaaosccn100C1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/04/2013 | 1530 | 10/19/2013 | 0100 | Missing |
Subject: | SGP/CCN/C1 - CCN shutdown for repair |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | A thermal electric cooler (TEC) was found to be not working. The instrument was shutdown until the TEC could be replaced. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2014 | 0614 | 07/21/2014 | 1340 | Does not affect quality |
Subject: | SGP/AOS/C1 - Frequent power and communication loss problems |
DataStreams: | sgpnoaaaosccn100C1.b1 |
Description: | There were frequent power and communication losses over the time period. The problem was resolved after power was redistributed and all data acquisition components were replaced. |
Measurements: | sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/12/2014 | 0614 | 07/15/2014 | 2250 | Does not affect quality |
Subject: | SGP/CCN/C1 - Frequent power and communication loss problems |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | There were frequent power and communication losses over the time period. The problem was resolved after power was redistributed and all data acquisition components were replaced. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/15/2014 | 2250 | 08/12/2014 | 1400 | Incorrect |
Subject: | SGP/CCN/C1 - CCN presure sensor failure |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | Sheath flow reading declined and couldn't be adjusted. Unit was shipped back to mentor for repair and calibration. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/30/2012 | 0000 | 06/11/2013 | 2359 | Incorrect |
Subject: | SGP/CCN/C1 - Incorrect supersaturation stepping method |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | The supersaturation stepping method over this period is not consistent with the ccnavg VAP expectations. Therefore, the averaging of the fields done by the CCNAVG VAP are incorrect - the output of the CCNAVG VAP for all fields is bad over this entire period. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/23/2014 | 1954 | 10/23/2014 | 2125 | Missing |
Subject: | SGP/AOS/C1 - New UPS installed |
DataStreams: | sgpnoaaaosccn100C1.b1 |
Description: | System down while new UPS is installed |
Measurements: | sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/15/2009 | 1600 | 08/03/2009 | 0000 | Suspect |
Subject: | SGP/CCN/C1 - CCN signal is low from possible leak and decline in laser power |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | CCN was returned with new laser. Signal was unusually low and flow calibration had changed substantially. A leak is suspect as well as declining laser power as new laser was defective. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/22/2007 | 2200 | 10/23/2007 | 1515 | Incorrect |
Subject: | SGP/CCN/C1 - Incorrect temperature gains or settings |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | New software was installed with instrument and temperature gains didn't transfer correctly. Super saturation values are incorrect and instrument tended to overheat in an attempt to reach set point values. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/11/2013 | 1926 | 04/08/2015 | 1730 | Suspect |
Subject: | SGP/CCN/C1 - Instrument under counting at lowest super saturation value |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | The slow temperature decay of the OPC induced droplet evaporation and caused an under counting of the droplets at the lowest %ss, 0.15. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2014 | 0700 | 03/15/2015 | 2045 | Missing |
Subject: | SGP/CCN/C1 - Removed for calibration |
DataStreams: | sgpaosccnavgC1.c1 |
Description: | Instrument was removed for calibration and scanning evaluation. |
Measurements: | sgpaosccnavgC1.c1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/20/2015 | 2044 | 10/24/2015 | 2300 | Incorrect |
Subject: | SGP/AOS/CCN/CLAP/CPC/NEPHELOMETER/PSAP/C1 - System Maintenance |
DataStreams: | sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1 |
Description: | The system was offline for cleaning and calibration. |
Measurements: | sgpaosccnavgC1.c1:
sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/07/2015 | 1552 | 12/08/2015 | 1649 | Incorrect |
Subject: | SGP/AOS/C1 - Pump and blower failure |
DataStreams: | sgpaosccnavgC1.c1, sgpnoaaaosccn100C1.b1 |
Description: | The fuse on the power strip with the pump and blower blew. The blower motor seized and the blower had to be replaced. |
Measurements: | sgpaosccnavgC1.c1:
sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/17/2015 | 0425 | 11/17/2015 | 1600 | Incorrect |
Subject: | SGP/AOS/C1 - Power outage |
DataStreams: | sgpnoaaaosccn100C1.b1 |
Description: | Power outage, no data collected. |
Measurements: | sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/26/2015 | 0000 | 01/01/2016 | 2359 | Does not affect quality |
Subject: | SGP Site Power Outage Due to Winter Storm |
DataStreams: | sgpnoaaaosccn100C1.b1, sgpaosccnavgC1.c1 |
Description: | On Saturday and Sunday, December 26 and 27th a powerful winter storm (Goliath) brought sleet and freezing rain to much of Oklahoma with wind gusts of 45 MPH at times. The result was very very slick roads and mass power outages. The SGP central facility lost power at 8:55 PM CDT on Sunday, December 27th. The power was restored on 7:56 PM CDT on Wednesday, December 30th Site Emergency procedures were activated and Technicians drained water out of instrument reservoirs and other measures as needed, early Monday morning. On Wednesdays as roads and power started was restored to much of the areas east of I-35 , Site Technicians performed maintenance on several Extended Facilities. During this period, data are potentially missing, incorrect, or suspect. Carefully inspect the data before use. |
Measurements: | sgpaosccnavgC1.c1:
sgpnoaaaosccn100C1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/18/2016 | 0400 | 03/18/2016 | 1100 | Does not affect quality |
03/25/2016 | 0100 | 03/25/2016 | 0700 | Does not affect quality |
Subject: | SGP/AOS/C1 - Wildfire smoke plumes advecting over facility |
DataStreams: | sgpnoaaaosccn100C1.b1 |
Description: | Smoke from nearby wildfires advected through SGP C1 during the listed affected time ranges, resulting in high signals observable in many of the listed AOS datastreams. |
Measurements: | sgpnoaaaosccn100C1.b1:
|