DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D160909.12 | SGP/ACSM/C1 - Filter switching valve not operating | sgpaosacsmC1.b1 | Incorrect |
D161005.1 | SGP/ACSM/C1 - Low inlet pressure in sample position | sgpaosacsmC1.b1 | Suspect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/12/2016 | 2055 | 08/31/2016 | 1249 | Incorrect |
Subject: | SGP/ACSM/C1 - Filter switching valve not operating |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Filter switching valve was not operating because of a faulty circuit board. Problem was fixed and data have been collected from the end of the specified time period. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/02/2016 | 1430 | 09/14/2016 | 1811 | Suspect |
Subject: | SGP/ACSM/C1 - Low inlet pressure in sample position |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The data are suspect because the new filter switching valve was out of adjustment causing low inlet pressure in the sample position. The problem was corrected by the onsite operator. |
Measurements: | sgpaosacsmC1.b1:
|