DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140109.1 | MAO/MET/M1 - ORG data incorrect | maometM1.b1 | Incorrect |
D140219.4 | MAO/MET/M1 - Reprocessing Notice (RID 1044) | maometM1.b1 | Does not affect quality |
D140228.2 | MAO/SONDE/M1 - Temperature Sensor Failure | maosondewnpnM1.b1 | Incorrect |
D140307.2 | MAO/CEIL/M1 - missing window blower | maoceilpblhtM1.a0 | Suspect |
D140314.4 | MAO/RWP/M1 - Instrument failure causing missing data | mao1290rwpwindconM1.a1 | Missing |
D140509.3 | MAO/RWP/M1 - Missing data due to power failure | mao1290rwpwindconM1.a1 | Missing |
D140529.1 | MAO/ECOR/M1 - CO2/H2O Means, CO2 flux, and Latent Heat Flux Incorrect | mao30ecorM1.b1 | Incorrect |
D140529.2 | MAO/ECOR/M1 - Wind Direction Too Low by About 10 Degrees. | mao30ecorM1.b1 | Incorrect |
D140530.1 | MAO/MET/M1 - Missing pressure data | maometM1.b1 | Missing |
D140530.3 | MAO/MET/M1 - Incorrect relative humidity, vapor pressure, and present weather detector data | maometM1.b1 | Incorrect |
D140625.1 | MAO/ECOR/M1 - Fluxes and wind measurements incorrect. | mao30ecorM1.b1 | Incorrect |
D140728.3 | MAO/MET/M1 - Optical rain gauge performance problems | maometM1.b1 | Suspect |
D140818.1 | MAO/RWP/M1 - Instrument failure | mao1290rwpwindconM1.a1 | Missing |
D140825.3 | MAO/ECOR/M1 - Latent Heat Flux and CO2 Flux Incorrect | mao30ecorM1.b1 | Incorrect |
D140902.5 | MAO/AOS/M1 - System down for maintenance | maonoaaaosavgM1.b1 | Incorrect |
D140905.1 | MAO/ECOR/M1 - Data Missing | mao30ecorM1.b1 | Missing |
D140929.1 | MAO/MET/M1 - Present weather detector power issue | maometM1.b1 | Incorrect |
D141002.1 | MAO/WACR/M1 - Instrument not operational | maowacrM1.a1 | Missing |
D141024.1 | MAO/WACR/M1 - Reprocessed and Corrected: Incorrect polarization status | maowacrM1.a1 | Does not affect quality |
D141030.3 | MAO/AOS/M1 - Sample line disconnected | maonoaaaosavgM1.b1 | Incorrect |
D141030.5 | MAO/AOS/M1 - Dilution dryer momentarily off | maonoaaaosavgM1.b1 | Incorrect |
D141111.4 | MAO/SONDE/M1 - Reprocessing Notice (RID 1253) | maosondewnpnM1.b1 | Does not affect quality |
D141114.3 | MAO/RWP/M1 - DSP communications lost between controller and Piraq boards | mao1290rwpwindconM1.a1 | Missing |
D141223.4 | MAO/WACR/M1 - Instrument failure. | maowacrM1.a1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2013 | 0000 | 01/08/2014 | 1727 | Incorrect |
Subject: | MAO/MET/M1 - ORG data incorrect |
DataStreams: | maometM1.b1 |
Description: | Optical rain gauge (ORG) precipitation values were incorrect the specified time due to a wiring problem. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2013 | 0000 | 02/19/2014 | 2359 | Does not affect quality |
Subject: | MAO/MET/M1 - Reprocessing Notice (RID 1044) |
DataStreams: | maometM1.b1 |
Description: | See below for details of reprocessing that was completed on 20140304. Original DQR category: 3 (Yellow, Suspect). ----------------------------------------------------------- The netcdf files split routinely due to a bug in the ingest. The ingest has been corrected and re-released to production. Data will be reprocessed. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/26/2014 | 2330 | 02/27/2014 | 0028 | Incorrect |
Subject: | MAO/SONDE/M1 - Temperature Sensor Failure |
DataStreams: | maosondewnpnM1.b1 |
Description: | It appears that the temperature sensor failed on release of balloon |
Measurements: | maosondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/21/2013 | 0451 | 03/03/2014 | 1659 | Suspect |
Subject: | MAO/CEIL/M1 - missing window blower |
DataStreams: | maoceilpblhtM1.a0 |
Description: | The window blower was missing from the VCEIL at deployment and could not be found. A new unit was ordered and installed. Data may be affected during periods of high humidity, condensation, and precipitation. |
Measurements: | maoceilpblhtM1.a0:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/15/2014 | 1900 | 02/24/2014 | 1500 | Missing |
01/01/2014 | 0000 | 02/14/2014 | 2019 | Missing |
Subject: | MAO/RWP/M1 - Instrument failure causing missing data |
DataStreams: | mao1290rwpwindconM1.a1 |
Description: | System under repair. No data available |
Measurements: | mao1290rwpwindconM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/13/2014 | 0613 | 04/13/2014 | 1809 | Missing |
Subject: | MAO/RWP/M1 - Missing data due to power failure |
DataStreams: | mao1290rwpwindconM1.a1 |
Description: | No data available |
Measurements: | mao1290rwpwindconM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/03/2014 | 1800 | 05/21/2014 | 1400 | Incorrect |
Subject: | MAO/ECOR/M1 - CO2/H2O Means, CO2 flux, and Latent Heat Flux Incorrect |
DataStreams: | mao30ecorM1.b1 |
Description: | All CO2 and water vapor means, CO2 flux, and latent heat flux are incorrect for the period. |
Measurements: | mao30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/03/2014 | 1800 | 05/22/2014 | 1530 | Incorrect |
Subject: | MAO/ECOR/M1 - Wind Direction Too Low by About 10 Degrees. |
DataStreams: | mao30ecorM1.b1 |
Description: | ECOR wind direction was about 10 degrees low during this period. |
Measurements: | mao30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/12/2014 | 1744 | 05/20/2014 | 1954 | Missing |
Subject: | MAO/MET/M1 - Missing pressure data |
DataStreams: | maometM1.b1 |
Description: | Atmospheric pressure data is missing due to power and wiring issues during this time. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/17/2014 | 0141 | 05/17/2014 | 1557 | Incorrect |
05/12/2014 | 1744 | 05/16/2014 | 1838 | Incorrect |
Subject: | MAO/MET/M1 - Incorrect relative humidity, vapor pressure, and present weather detector |
DataStreams: | maometM1.b1 |
Description: | The relative humidity, vapor pressure, and all data from the PWD are incorrect due to power and wiring problems during this time. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/21/2014 | 0130 | 06/21/2014 | 1400 | Incorrect |
Subject: | MAO/ECOR/M1 - Fluxes and wind measurements incorrect. |
DataStreams: | mao30ecorM1.b1 |
Description: | The measurements were incorrect because the ECOR tripod was tipped over. |
Measurements: | mao30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/20/2013 | 0000 | 10/14/2014 | 2118 | Suspect |
Subject: | MAO/MET/M1 - Optical rain gauge performance problems |
DataStreams: | maometM1.b1 |
Description: | Condensation inside the optical rain gauge lenses caused false precipitation to be recorded. Sensor replacements also showed signs of false precipitation. User discretion advised. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/05/2014 | 2310 | 08/09/2014 | 1556 | Missing |
Subject: | MAO/RWP/M1 - Instrument failure |
DataStreams: | mao1290rwpwindconM1.a1 |
Description: | No data available |
Measurements: | mao1290rwpwindconM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/21/2014 | 0130 | 08/24/2014 | 1500 | Incorrect |
Subject: | MAO/ECOR/M1 - Latent Heat Flux and CO2 Flux Incorrect |
DataStreams: | mao30ecorM1.b1 |
Description: | The analog outputs of the H2O/CO2 analyzer were not working; therefore latent heat flux and CO2 flux were incorrect during the specified period. |
Measurements: | mao30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/14/2014 | 1200 | 08/20/2014 | 1400 | Incorrect |
Subject: | MAO/AOS/M1 - System down for maintenance |
DataStreams: | maonoaaaosavgM1.b1 |
Description: | Instruments were periodically offline or sampling room air during the cleaning, repair and calibrations. |
Measurements: | maonoaaaosavgM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
08/27/2014 | 1200 | 08/28/2014 | 2030 | Missing |
Subject: | MAO/ECOR/M1 - Data Missing |
DataStreams: | mao30ecorM1.b1 |
Description: | All data was missing during then stated period. |
Measurements: | mao30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/17/2014 | 1530 | 09/28/2014 | 1600 | Incorrect |
Subject: | MAO/MET/M1 - Present weather detector power issue |
DataStreams: | maometM1.b1 |
Description: | A broken power connection caused incorrect data from the present weather detector during this time. |
Measurements: | maometM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/13/2014 | 1700 | 09/29/2014 | 2106 | Missing |
Subject: | MAO/WACR/M1 - Instrument not operational |
DataStreams: | maowacrM1.a1 |
Description: | Instrument went down and needed to be restarted. Data is missing and not recoverable. |
Measurements: | maowacrM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/18/2014 | 1350 | 04/28/2014 | 1739 | Does not affect quality |
Subject: | MAO/WACR/M1 - Reprocessed and Corrected: Incorrect polarization status |
DataStreams: | maowacrM1.a1 |
Description: | DATA REPROCESSED and CORRECTED: During this period, raw data was incorrectly identifying polarization status for hour-long periods, intermittently. So some hour-long periods were correct, and some hour-long periods had polarization status swapped. The ingest was modified to identify the good and bad hourly periods from reflectivity and linear depolarization ratio images. A new ingest software was released and the WACR data were reprocessed. The data were Archived in July 2016. Please retrieve the reprocessed data for use in your analysis. You may find these data using the ARM Discovery tool at the following link: http://www.archive.arm.gov/discovery/#v/results/s/finst::wacr/fsite::mao/ffac::mao.M1 ORIGINAL DQR CATEGORY: Red( Incorrect) REPROCESSED and CORRECTED DATA ARCHIVED:2016-07 REPROCESS IDENTIFICATION NUMBER: RID 1249 --------------------------------------------------- --- ORIGINAL DQR DESCRIPTION --- --------------------------------------------------- During this period, raw data was incorrectly identifying polarization status for hour-long periods, intermittently. So some hour-long periods are correct, and some hour-long periods have polarization status swapped. |
Measurements: | maowacrM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/09/2014 | 1600 | 10/10/2014 | 1635 | Incorrect |
Subject: | MAO/AOS/M1 - Sample line disconnected |
DataStreams: | maonoaaaosavgM1.b1 |
Description: | While replacing the flow meter the tech left part of the sample line disconnected the system was sampling room air. |
Measurements: | maonoaaaosavgM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/16/2014 | 1309 | 10/16/2014 | 1330 | Incorrect |
Subject: | MAO/AOS/M1 - Dilution dryer momentarily off |
DataStreams: | maonoaaaosavgM1.b1 |
Description: | |
Measurements: | maonoaaaosavgM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/17/2014 | 0528 | 10/17/2014 | 0718 | Does not affect quality |
Subject: | MAO/SONDE/M1 - Reprocessing Notice (RID 1253) |
DataStreams: | maosondewnpnM1.b1 |
Description: | See below for details of reprocessing that was completed on 20150121. Original DQR category: 4(Red, Incorrect) ------------------------------------------------------------------------------- The operator incorrectly entered the surface RH value as 10% when it was actually 100%. |
Measurements: | maosondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/31/2014 | 1000 | 10/31/2014 | 1200 | Missing |
11/01/2014 | 1300 | 11/01/2014 | 1400 | Missing |
10/30/2014 | 1700 | 10/30/2014 | 1800 | Missing |
10/31/2014 | 1500 | 10/31/2014 | 1700 | Missing |
11/01/2014 | 1600 | 11/01/2014 | 1830 | Missing |
11/03/2014 | 0630 | 11/05/2014 | 1600 | Missing |
10/31/2014 | 1800 | 10/31/2014 | 2100 | Missing |
11/02/2014 | 1600 | 11/03/2014 | 0600 | Missing |
11/01/2014 | 1000 | 11/01/2014 | 1200 | Missing |
11/01/2014 | 1945 | 11/01/2014 | 2015 | Missing |
11/02/2014 | 0005 | 11/02/2014 | 1330 | Missing |
Subject: | MAO/RWP/M1 - DSP communications lost between controller and Piraq boards |
DataStreams: | mao1290rwpwindconM1.a1 |
Description: | No data available because no transmit or receive functionality |
Measurements: | mao1290rwpwindconM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/16/2014 | 0012 | 04/16/2014 | 0212 | Incorrect |
04/16/2014 | 0840 | 04/16/2014 | 1616 | Incorrect |
04/16/2014 | 0620 | 04/16/2014 | 0714 | Incorrect |
Subject: | MAO/WACR/M1 - Instrument failure. |
DataStreams: | maowacrM1.a1 |
Description: | The moments data are incorrect. |
Measurements: | maowacrM1.a1:
|