D140611.1ENA/PSAP/C1 - Reprocess: Ingest problemenaaospsap3wC1.b1Incorrect
D140618.1ENA/PSAP/C1 - Electronics failureenaaospsap3wC1.b1Incorrect
D140618.2ENA/CPC/NEPH/PSAP/C1 - Impactor not installedenaaospsap3wC1.b1Does not affect quality
D140618.3ENA/PSAP/C1 - Board drop outs in PSAPenaaospsap3wC1.b1Does not affect quality
D140820.2ENA/PSAP/C1 - Data processing problemenaaospsap3wC1.b1Incorrect
D140926.2ENA/PSAP/C1 - Computer Failureenaaospsap3wC1.b1Missing
D141110.1ENA/PSAP/C1 - Incorrect instrument configurationenaaospsap3wC1.b1Missing
D150220.7ENA/PSAP/C1 - Pump failure resulted in incorrect dataenaaospsap3wC1.b1Incorrect
D150506.2ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visitenaaospsap3wC1.b1Incorrect
D150511.6ENA/PSAP/C1 - Data Gaps > 1h due to Tr<0.7, pump turned offenaaospsap3wC1.b1Incorrect
D150701.8ENA/PSAP/C1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1
level
enaaospsap3wC1.b1Does not affect quality
D151106.2ENA/NEPHELOMETER/PSAP/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor stuck at 10 micronsenaaospsap3wC1.b1Does not affect quality
D151112.5ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switchingenaaospsap3wC1.b1Does not affect quality
D151112.6ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connectedenaaospsap3wC1.b1Does not affect quality
D160427.7ENA/C1 - Reprocess: Slightly incorrect location fieldsenaaospsap3wC1.b1Suspect
D160613.1ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switchingenaaospsap3w1sC1.b1, enaaospsap3wC1.b1Suspect
D160706.1ENA/PSAP/C1 - Noisy green channelenaaospsap3w1sC1.b1, enaaospsap3wC1.b1Suspect
D161025.6ENA/AOS/C1 - Instrument mentors working at the siteenaaospsap3wC1.b1Suspect
D170303.12ENA/AOS/C1 - Instrumentenaaospsap3wC1.b1Does not affect quality


DQRID : D140611.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/04/2013140804/23/20142359Incorrect
Subject:
ENA/PSAP/C1 - Reprocess: Ingest problem
DataStreams:enaaospsap3wC1.b1
Description:
A logic error in the code cause the program to fail when there was no impactor data 
available.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D140618.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/31/2014111602/01/20140000Incorrect
01/18/2014091501/29/20140000Incorrect
Subject:
ENA/PSAP/C1 - Electronics failure
DataStreams:enaaospsap3wC1.b1
Description:
Electronics failure causing incorrect values.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D140618.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/01/2013000002/19/20141351Does not affect quality
Subject:
ENA/CPC/NEPH/PSAP/C1 - Impactor not installed
DataStreams:enaaospsap3wC1.b1
Description:
No impactor installed.  All data is the same as when impactor at 10 um setting.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D140618.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/01/2014000002/22/20140000Does not affect quality
Subject:
ENA/PSAP/C1 - Board drop outs in PSAP
DataStreams:enaaospsap3wC1.b1
Description:
Board seating appears to have caused intermittent drop outs.  These ceased after Mentor 
visit and reseating of all boards.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D140820.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/19/2014035103/11/20141500Incorrect
Subject:
ENA/PSAP/C1 - Data processing problem
DataStreams:enaaospsap3wC1.b1
Description:
The impactor data was not available during this period and the impactor_setting variable 
in these datastreams shows a constant value of 10 um.  The impactor data has since been 
recovered.
Measurements:enaaospsap3wC1.b1:
  • impactor_setting


Back To Table of Contents

DQRID : D140926.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/20/2014113209/22/20141548Missing
Subject:
ENA/PSAP/C1 - Computer Failure
DataStreams:enaaospsap3wC1.b1
Description:
Data were lost due to a failed CPU unit
Measurements:enaaospsap3wC1.b1:
  • North latitude(lat)
  • Altitude above mean sea level(alt)
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • East longitude(lon)
  • transmittance_B
  • base_time
  • time_offset
  • time
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • dilution_flow_valve_position
  • dilution_flow_setpoint
  • transmittance_R


Back To Table of Contents

DQRID : D141110.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/11/2014142910/15/20140754Missing
Subject:
ENA/PSAP/C1 - Incorrect instrument configuration
DataStreams:enaaospsap3wC1.b1
Description:
Instrument set to incorrect output configuration for all serial output datastreams 
resulting in no data output.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D150220.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/23/2015203801/24/20151215Incorrect
01/21/2015111201/22/20151022Incorrect
Subject:
ENA/PSAP/C1 - Pump failure resulted in incorrect data
DataStreams:enaaospsap3wC1.b1
Description:
Data incorrect as a result of a pump failure.
Measurements:enaaospsap3wC1.b1:
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)


Back To Table of Contents

DQRID : D150506.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/22/2015170004/27/20151700Incorrect
Subject:
ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit
DataStreams:enaaospsap3wC1.b1
Description:
There was visit of 7 AOS instrument mentors to the ENA AOS during this period of 
instrument repair, calibration and modification. Instruments were occasionally off, or with no 
sample flow, or receiving calibration samples, etc. There was no productive sampling of 
ambient air during this period other than for testing. No data should be used for this period 
as ambient samples.

Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas 
instrument.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D150511.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/15/2015011603/15/20150915Incorrect
03/14/2015064103/14/20150954Incorrect
03/27/2015003103/27/20150250Incorrect
03/28/2015024603/28/20150459Incorrect
03/27/2015105203/27/20151424Incorrect
03/27/2015025003/27/20150714Incorrect
Subject:
ENA/PSAP/C1 - Data Gaps > 1h due to Tr<0.7, pump turned off
DataStreams:enaaospsap3wC1.b1
Description:
Vendor specifies Transmission greater than 0.7 for data analysis
Measurements:enaaospsap3wC1.b1:
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)


Back To Table of Contents

DQRID : D150701.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/04/2013000010/01/20152359Does not affect quality
Subject:
ENA/PSAP/C1 - Reprocess: reprocessing from raw data in preparation for not-yet-released b1 
level
DataStreams:enaaospsap3wC1.b1
Description:
End to end reprocessing to generate new a1 datastreams is needed so that when the 
harmonization software is released to production, we will be ready with the a1 datastreams that 
are input to the subsequent processing software.
Measurements:enaaospsap3wC1.b1:
  • North latitude(lat)
  • Altitude above mean sea level(alt)
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • East longitude(lon)
  • transmittance_B
  • base_time
  • time_offset
  • time
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • dilution_flow_valve_position
  • dilution_flow_setpoint
  • transmittance_R


Back To Table of Contents

DQRID : D151106.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/02/2015160711/03/20151409Does not affect quality
Subject:
ENA/NEPHELOMETER/PSAP/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor stuck at 10 microns
DataStreams:enaaospsap3wC1.b1
Description:
Upstream impactor was stuck at the 10 micron position. Data for the period is valid but 
representative of an effective 10 micrometer impactor setting only.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D151112.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/27/2015105203/30/20151800Does not affect quality
Subject:
ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor not switching
DataStreams:enaaospsap3wC1.b1
Description:
AOS tech was working in the AOS and accidentally unplugged the impactor power supply. This 
left the impactor in the 10 micron state.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D151112.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/19/2014135104/26/20151016Does not affect quality
Subject:
ENA/CAPS-PMEX/CPC/NEPHELOMETER/PSAP/C1 - Impactor size switch not connected
DataStreams:enaaospsap3wC1.b1
Description:
Impactor installed and plumbed, HOWEVER, the impactor's 'size read' switch NOT connected.
Measurements:enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • mass_flow
  • dilution_flow
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • transmittance_B
  • dilution_flow_valve_position
  • transmittance_R
  • dilution_flow_setpoint


Back To Table of Contents

DQRID : D160427.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/29/2013094104/30/20162359Suspect
Subject:
ENA/C1 - Reprocess: Slightly incorrect location fields
DataStreams:enaaospsap3wC1.b1
Description:
Some of the instruments are being processed with the previous "GRW-Azores" lat/lon/alt 
instead of the new "ENA-Azores" lat/lon/alt. The differences are minor, but will be 
corrected. The respective GRW and ENA lat, lon, and alt values are as follows:
 
GRW  lat = 39.0911 ;
ENA lat = 39.0916 ;
GRW  lon = -28.0297 ;
ENA  lon = -28.0257 ;
GRW  alt = 26 ;
ENA  alt = 30.48 ;
Measurements:enaaospsap3wC1.b1:
  • North latitude(lat)
  • Altitude above mean sea level(alt)
  • East longitude(lon)


Back To Table of Contents

DQRID : D160613.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/10/2016150006/12/20161100Suspect
Subject:
ENA/CAPS-PMEX/NEPHELOMETER/PSAP/C1 - Impactor not switching
DataStreams:enaaospsap3w1sC1.b1, enaaospsap3wC1.b1
Description:
Impactor power accidentally unplugged. Impactor not switching and stuck in 10um position. 
The quality of data variables is not affected.
Measurements:enaaospsap3w1sC1.b1:
  • Impactor particle size cut(impactor_state)

enaaospsap3wC1.b1:
  • Impactor particle size cut(impactor_state)
  • impactor_setting


Back To Table of Contents

DQRID : D160706.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/02/2016002404/19/20161535Suspect
Subject:
ENA/PSAP/C1 - Noisy green channel
DataStreams:enaaospsap3w1sC1.b1, enaaospsap3wC1.b1
Description:
Instrument's green channel became noisy as characterized by +/- excursions.  The green 
data should be used with caution.  Problem solved by reseating the boards.
Measurements:enaaospsap3w1sC1.b1:
  • Aerosol light absorption coefficient, green channel at dry or reference RH(Ba_G_Weiss)
  • Uncorrected aerosol light absorption coefficient, green channel at dry or
    reference RH(Ba_G_raw)
  • Transmittance, green channel(transmittance_green)

enaaospsap3wC1.b1:
  • transmittance_G
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • Transmittance, green channel(transmittance_green)
  • transmittance_green_raw
  • transmittance_green_normalization_factor


Back To Table of Contents

DQRID : D161025.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/26/2016083010/01/20160100Suspect
Subject:
ENA/AOS/C1 - Instrument mentors working at the site
DataStreams:enaaospsap3wC1.b1
Description:
Instrument mentors working at the site. Sample lines and electrical connections being 
reconfigured. Instruments that had no work done on them may still be affected.
Measurements:enaaospsap3wC1.b1:
  • tr_red
  • Light absorbance, green channel(absorbance_green)
  • transmittance_blue_normalization_factor
  • tr_green
  • transmittance_green_raw
  • transmittance_blue_raw
  • dilution_flow
  • dilution_correction_factor
  • Transmittance, green channel(transmittance_green)
  • transmittance_red_raw
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • dilution_flow_setpoint
  • spot_size_area
  • transmittance_G
  • sample_flow_rate
  • transmittance_red_normalization_factor
  • tr_blue
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • transmittance_B
  • Impactor particle size cut(impactor_state)
  • Transmittance, blue channel(transmittance_blue)
  • Light absorbance, blue channel(absorbance_blue)
  • mass_flow
  • Transmittance, red channel(transmittance_red)
  • filter_unstable
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)
  • impactor_setting
  • seconds_after_transition
  • dilution_flow_valve_position
  • Light absorbance, red channel(absorbance_red)
  • transmittance_green_normalization_factor
  • transmittance_R


Back To Table of Contents

DQRID : D170303.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/20/2016000003/03/20170000Does not affect quality
Subject:
ENA/AOS/C1 - Instrument
DataStreams:enaaospsap3wC1.b1
Description:
Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and 
the HT-DMA. Dry air supply for those instruments partially replaced by regular undried 
compressed air, lowering the delivery pressure. No direct effect on other instruments seen.
Measurements:enaaospsap3wC1.b1:
  • tr_red
  • Altitude above mean sea level(alt)
  • Light absorbance, green channel(absorbance_green)
  • transmittance_blue_normalization_factor
  • qc_impactor_state
  • tr_green
  • transmittance_green_raw
  • base_time
  • transmittance_blue_raw
  • time
  • qc_sample_flow_rate
  • dilution_flow
  • dilution_correction_factor
  • Transmittance, green channel(transmittance_green)
  • transmittance_red_raw
  • dilution_flow_setpoint
  • North latitude(lat)
  • spot_size_area
  • qc_transmittance_blue
  • sample_flow_rate
  • transmittance_red_normalization_factor
  • tr_blue
  • East longitude(lon)
  • Impactor particle size cut(impactor_state)
  • qc_transmittance_green
  • Transmittance, blue channel(transmittance_blue)
  • Light absorbance, blue channel(absorbance_blue)
  • time_offset
  • Transmittance, red channel(transmittance_red)
  • qc_transmittance_red
  • filter_unstable
  • seconds_after_transition
  • dilution_flow_valve_position
  • Light absorbance, red channel(absorbance_red)
  • transmittance_green_normalization_factor


Back To Table of Contents