Data Quality Reports for Session: 184329 User: gallof1 Completed: 03/17/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D150506.2ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site VisitenaaoscoC1.a1Incorrect
D151029.3ENA/TRACEGAS/C1 - Calibration turned offenaaoscoC1.a1Does not affect quality
D151029.4ENA/TRACEGAS/C1 - Noise in signalenaaoscoC1.a1Suspect
D160202.1ENA/TRACEGAS/C1 - Reduced calibration from lost instrument controlenaaoscoC1.a1Does not affect quality
D160630.1ENA/TRACEGAS/C1 - Data drop outenaaoscoC1.a1Incorrect
D160907.11ENA/TRACEGAS/C1 - Flow meters not operatingenaaoscoC1.a1Does not affect quality
D161025.6ENA/AOS/C1 - Instrument mentors working at the siteenaaoscoC1.a1Suspect
D161026.2ENA/AOSMET/OZONE/TRACEGAS/C1 - missing dataenaaoscoC1.a1Missing
D161216.5ENA/TRACEGAS/C1 - Power failureenaaoscoC1.a1Incorrect
D170303.12ENA/AOS/C1 - InstrumentenaaoscoC1.a1Does not affect quality


DQRID : D150506.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/22/2015170004/27/20151700Incorrect
Subject:
ENA/AOS/C1 - AOS Instrument Repair, Calibration and Modification: Mentor Site Visit
DataStreams:enaaoscoC1.a1
Description:
There was visit of 7 AOS instrument mentors to the ENA AOS during this period of 
instrument repair, calibration and modification. Instruments were occasionally off, or with no 
sample flow, or receiving calibration samples, etc. There was no productive sampling of 
ambient air during this period other than for testing. No data should be used for this period 
as ambient samples.

Note: Two datastream are missing. The CAPS Extinction Monitor and the Greenhouse gas 
instrument.
Measurements:enaaoscoC1.a1:
  • gas_pressure
  • ambient_temperature
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D151029.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/11/2015000007/14/20152359Does not affect quality
07/18/2015000007/21/20152359Does not affect quality
Subject:
ENA/TRACEGAS/C1 - Calibration turned off
DataStreams:enaaoscoC1.a1
Description:
Calibrations before and after period are very consistent. Data appears to be unaffected.
Measurements:enaaoscoC1.a1:
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)


Back To Table of Contents

DQRID : D151029.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/15/2015225507/19/20151924Suspect
Subject:
ENA/TRACEGAS/C1 - Noise in signal
DataStreams:enaaoscoC1.a1
Description:
Laser lock may have been lost.  Signal noisy on H2O and N2O channels
Measurements:enaaoscoC1.a1:
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D160202.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/25/2016174302/02/20161931Does not affect quality
Subject:
ENA/TRACEGAS/C1 - Reduced calibration from lost instrument control
DataStreams:enaaoscoC1.a1
Description:
Instrument control of calibration Mass Flow Controllers ceased. Unclear why. Lack of 
calibration does not significantly compromise performance.
Measurements:enaaoscoC1.a1:
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)


Back To Table of Contents

DQRID : D160630.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/11/2016170403/11/20161836Incorrect
Subject:
ENA/TRACEGAS/C1 - Data drop out
DataStreams:enaaoscoC1.a1
Description:
Instrument data drop out for an unknown reason.  No valid data.
Measurements:enaaoscoC1.a1:
  • mass_flow_through_MFC_2
  • gas_pressure
  • ambient_temperature
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • set_point_for_MFC_1
  • Calculated CO concentration(co)
  • N2O mole fraction of during batch measurement(n2o)
  • mass_flow_through_MFC_1
  • set_point_for_MFC_2
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • valve_position_MFC_2
  • valve_position_MFC_1
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D160907.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/25/2016114109/07/20161900Does not affect quality
Subject:
ENA/TRACEGAS/C1 - Flow meters not operating
DataStreams:enaaoscoC1.a1
Description:
Flow meters were not operating so that there were no standard adds for calibration.
Measurements:enaaoscoC1.a1:
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)


Back To Table of Contents

DQRID : D161025.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/26/2016083010/01/20160100Suspect
Subject:
ENA/AOS/C1 - Instrument mentors working at the site
DataStreams:enaaoscoC1.a1
Description:
Instrument mentors working at the site. Sample lines and electrical connections being 
reconfigured. Instruments that had no work done on them may still be affected.
Measurements:enaaoscoC1.a1:
  • mass_flow_through_MFC_2
  • gas_pressure
  • ambient_temperature
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • set_point_for_MFC_1
  • Calculated CO concentration(co)
  • N2O mole fraction of during batch measurement(n2o)
  • mass_flow_through_MFC_1
  • set_point_for_MFC_2
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • valve_position_MFC_2
  • valve_position_MFC_1
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D161026.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/25/2016005010/25/20160518Missing
Subject:
ENA/AOSMET/OZONE/TRACEGAS/C1 - missing data
DataStreams:enaaoscoC1.a1
Description:
The Unit 1 computer was found stuck in boot screen and had to be powered off to reboot.  
Data is missing for this period.
Measurements:enaaoscoC1.a1:
  • mass_flow_through_MFC_2
  • Altitude above mean sea level(alt)
  • gas_pressure
  • base_time
  • time_offset
  • ambient_temperature
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • East longitude(lon)
  • set_point_for_MFC_1
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)
  • mass_flow_through_MFC_1
  • set_point_for_MFC_2
  • time
  • North latitude(lat)
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • valve_position_MFC_2
  • valve_position_MFC_1
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D161216.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/11/2016161011/12/20160937Incorrect
Subject:
ENA/TRACEGAS/C1 - Power failure
DataStreams:enaaoscoC1.a1
Description:
Instrument restarted after a power failure. Data not correct.  Problem resolved after 
Operator intervention.
Measurements:enaaoscoC1.a1:
  • mass_flow_through_MFC_2
  • gas_pressure
  • ambient_temperature
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • set_point_for_MFC_1
  • Calculated CO concentration(co)
  • N2O mole fraction of during batch measurement(n2o)
  • mass_flow_through_MFC_1
  • set_point_for_MFC_2
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • valve_position_MFC_2
  • valve_position_MFC_1
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents

DQRID : D170303.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/20/2016000003/03/20170000Does not affect quality
Subject:
ENA/AOS/C1 - Instrument
DataStreams:enaaoscoC1.a1
Description:
Pentras dry air supply intermittently not working. Dry air missing for the humidigraph and 
the HT-DMA. Dry air supply for those instruments partially replaced by regular undried 
compressed air, lowering the delivery pressure. No direct effect on other instruments seen.
Measurements:enaaoscoC1.a1:
  • mass_flow_through_MFC_2
  • Altitude above mean sea level(alt)
  • gas_pressure
  • base_time
  • time_offset
  • ambient_temperature
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • East longitude(lon)
  • set_point_for_MFC_1
  • N2O mole fraction of during batch measurement(n2o)
  • Calculated CO concentration(co)
  • mass_flow_through_MFC_1
  • set_point_for_MFC_2
  • time
  • North latitude(lat)
  • gas_temperature
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • valve_position_MFC_2
  • valve_position_MFC_1
  • Water vapor mixing ratio calculated with nominal sensitivity correction(h2o)


Back To Table of Contents



END OF DATA