DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D061214.1 | TWP/SKYRAD/C3 - NIP instrument failure | twpskyrad60sC3.b1 | Incorrect |
D080414.1 | TWP/SKYRAD/C3 - IRT power supply | twpskyrad60sC3.b1 | Incorrect |
D080904.6 | TWP/SKYRAD/C3 - Contamination | twpskyrad60sC3.b1 | Suspect |
D081222.2 | TWP/SKYRAD/C3 - Direct normal instrument failure | twpskyrad60sC3.b1 | Incorrect |
D090129.1 | TWP/SKYRAD/C3 - Tracker inoperative | twpskyrad60sC3.b1 | Incorrect |
D091112.6 | TWP/SKYRAD/C3 - IRT power supply | twpskyrad60sC3.b1 | Missing |
D100706.1 | TWP/SKYRAD/C3 - IRT power supply | twpskyrad60sC3.b1 | Missing |
D131220.3 | TWP/SKYRAD/C3 - IRT bias | twpskyrad60sC3.b1 | Suspect |
D141204.1 | TWP/SKYRAD/C3 - Reprocess: Incorrect qc_flag values | twpskyrad60sC3.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/13/2006 | 0342 | 01/16/2007 | 0440 | Incorrect |
Subject: | TWP/SKYRAD/C3 - NIP instrument failure |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The NIP (Short Direct Normal) was replaced after it displayed symptoms of failure associated with either unknown precipitation related issues or a possible lightning strike. NIP data are not recoverable. Direct normal data can be reconstructed from global hemispheric and global diffuse data using the following relationship: global hemispheric = direct*cos(sza) + diffuse hemispheric |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/11/2008 | 1901 | 04/14/2008 | 2205 | Incorrect |
Subject: | TWP/SKYRAD/C3 - IRT power supply |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The 12-to-24V DC converter the provides power to the IRT had shut down and was restarted by power-cycling. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/08/2008 | 0000 | 09/05/2008 | 0053 | Suspect |
Subject: | TWP/SKYRAD/C3 - Contamination |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The IRT exhibited a positive bias compared to AERI. The problem was corrected when the lens and mirror were cleaned. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/14/2008 | 0457 | 12/30/2008 | 0040 | Incorrect |
Subject: | TWP/SKYRAD/C3 - Direct normal instrument failure |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The NIP instrument (short_direct_normal data stream) failed giving bad data during this time period. The failure start time coincides with a reported DC power failure but no connection between the reported power failure and the failure of this instrument could be determined. Ops also reported the tracker clock running 2.9 minutes slow during maintenance on 12/22 0300GMT (see CS-2008-0714). Again, the slow clock is not suspect and did not appear to affect alignment. Ops replaced the NIP instrument on 12/30 and short_direct_normal data resumed normal operation. Disregard the short_direct_normal (NIP) data during this time period or calculate using the global and diffuse instruments and the following relationship: short_direct_normal = ((down_short_hemisp - down_short_diffuse_hemisp))/cos(zenith angle) |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/23/2009 | 2100 | 01/27/2009 | 0715 | Incorrect |
Subject: | TWP/SKYRAD/C3 - Tracker inoperative |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The tracker at Darwin failed 01/23. The tracker was power cycled on 01/27 and it returned to normal operation. The affected data streams are short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48). The down_short_hemisp are not affected. The down_long_hemisp1 and 2 (PIR1 and PIR2) should not be affected but direct sun can affect individual instrument performance. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/07/2009 | 1940 | 11/11/2009 | 0100 | Missing |
Subject: | TWP/SKYRAD/C3 - IRT power supply |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The IRT external DC power supply had shut down for an unknown reason. The problem was corrected by restarting the power supply. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/04/2010 | 0929 | 05/05/2010 | 2345 | Missing |
Subject: | TWP/SKYRAD/C3 - IRT power supply |
DataStreams: | twpskyrad60sC3.b1 |
Description: | The IRT power supply failed and was replaced. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/25/2002 | 1200 | 06/29/2007 | 0001 | Suspect |
Subject: | TWP/SKYRAD/C3 - IRT bias |
DataStreams: | twpskyrad60sC3.b1 |
Description: | Measurements of sky temperature, especially those <250K, were positively biased compared with AERI because there was no compensation for the reflectivity of the gold mirror. The problem was corrected by changing the IRT emissivity from 1.000 to 0.987. Use sky temperature measurements less than 230K with caution. Biases as large as 30K have been observed when the actual temperature, as measured by AERI, is less than 210 K. |
Measurements: | twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/28/2013 | 2345 | 12/08/2014 | 2359 | Does not affect quality |
Subject: | TWP/SKYRAD/C3 - Reprocess: Incorrect qc_flag values |
DataStreams: | twpskyrad60sC3.b1 |
Description: | When the IRTs were upgraded, the values of sky_ir_temp:min and sky_ir_temp:max were not changed. The quality control fields for these limits are incorrect. |
Measurements: | twpskyrad60sC3.b1:
|