DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D100129.1 | GRW/AOS/M1 - CPC TEC failure | grwaosM1.a1 | Missing |
D101020.2 | GRW/AOS/M1 - Data processing problem | grwaosM1.a1 | Does not affect quality |
D120820.1 | GRW/AOS/M1 - CPC undercounting | grwaosM1.a1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/19/2009 | 1844 | 01/15/2010 | 1820 | Missing |
Subject: | GRW/AOS/M1 - CPC TEC failure |
DataStreams: | grwaosM1.a1 |
Description: | The thermal electric cooler, which controls the temperature in the Condensation Particle Counter, failed. A new TEC was mailed to GRW. The TEC was replaced and the CPC was working again on Jan. 15th, 2010. |
Measurements: | grwaosM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/15/2009 | 0000 | 12/31/2009 | 0000 | Does not affect quality |
Subject: | GRW/AOS/M1 - Data processing problem |
DataStreams: | grwaosM1.a1 |
Description: | The AOS data was reprocessed for the AMF, SGP and NSA sites. There was an error in the processing of the corrections for the PSAP absorption signal such that a low pass filter was implemented for the scattering data used in these corrections.The difference between the corrected and previous data is < 0.05 Mm-1 (below the instrument detection limit) and should not have a significant effect on the previous data. Data will be reprocessed at a later time when more significant changes to the data processing are implemented. |
Measurements: | grwaosM1.a1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/27/2009 | 0000 | 09/08/2009 | 0825 | Incorrect |
Subject: | GRW/AOS/M1 - CPC undercounting |
DataStreams: | grwaosM1.a1 |
Description: | The CPC was undercounting by about 10-20 percent relative to the CCN. After cleaning the CPC and replacing the butanol, the signal increased to above that of the CCN. Increase CPC signal so that it equals the CCN signal at the maximum SS value. The corrected signal will represent relative and not an absolute particle concentration. |
Measurements: | grwaosM1.a1:
|