D160909.12SGP/ACSM/C1 - Filter switching valve not operatingsgpaosacsmC1.b1, sgpaosacsmC1.a1Incorrect
D161005.1SGP/ACSM/C1 - Low inlet pressure in sample positionsgpaosacsmC1.b1, sgpaosacsmC1.a1Suspect


DQRID : D160909.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/12/2016205508/31/20161249Incorrect
Subject:
SGP/ACSM/C1 - Filter switching valve not operating
DataStreams:sgpaosacsmC1.b1, sgpaosacsmC1.a1
Description:
Filter switching valve was not operating because of a faulty circuit board.  Problem was 
fixed and data have been collected from the end of the specified time period.
Measurements:sgpaosacsmC1.a1:
  • flow_rate
  • vaporizer
  • inlet_closed
  • Mass concentration of chloride, ambient aerosol in air(chloride)
  • emission_current
  • Mass concentration of total organics, ambient aerosol in air.(total_organics)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • inlet_open
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of nitrate, ambient aerosol in air.(nitrate)

sgpaosacsmC1.b1:
  • org_mx_err
  • Mass concentration of nitrate, ambient aerosol in air.(nitrate)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of chloride, ambient aerosol in air(chloride)
  • org_mx
  • Mass concentration of total organics, ambient aerosol in air.(total_organics)
  • amus
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)


Back To Table of Contents

DQRID : D161005.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/02/2016143009/14/20161811Suspect
Subject:
SGP/ACSM/C1 - Low inlet pressure in sample position
DataStreams:sgpaosacsmC1.b1, sgpaosacsmC1.a1
Description:
The data are suspect because the new filter switching valve was out of adjustment causing 
low inlet pressure in the sample position.  The problem was corrected by the onsite 
operator.
Measurements:sgpaosacsmC1.a1:
  • flow_rate
  • vaporizer
  • inlet_closed
  • Mass concentration of chloride, ambient aerosol in air(chloride)
  • emission_current
  • Mass concentration of total organics, ambient aerosol in air.(total_organics)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • inlet_open
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of nitrate, ambient aerosol in air.(nitrate)

sgpaosacsmC1.b1:
  • org_mx_err
  • Mass concentration of nitrate, ambient aerosol in air.(nitrate)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of chloride, ambient aerosol in air(chloride)
  • org_mx
  • Mass concentration of total organics, ambient aerosol in air.(total_organics)
  • amus
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)