Data Quality Reports for Session: 184488 User: moffettc1 Completed: 03/22/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D121201.2NSA/AMC/C1 - Single sensor (vwc_4) failurensaamcC1.b1Incorrect
D160212.20NSA/AMC/C1 - Instrument probationary periodnsaamcC1.b1Suspect


DQRID : D121201.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/13/2012170011/05/20122100Incorrect
Subject:
NSA/AMC/C1 - Single sensor (vwc_4) failure
DataStreams:nsaamcC1.b1
Description:
The vwc_4 (volumetric water content 4) sensor malfunctioned on 8/13.  It was 
intermittently bad throughout the specified period.
Measurements:nsaamcC1.b1:
  • Average volumetric water content sensor 4(vwc_4)
  • vwc_4_std


Back To Table of Contents

DQRID : D160212.20
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/08/2012020012/06/20141659Suspect
Subject:
NSA/AMC/C1 - Instrument probationary period
DataStreams:nsaamcC1.b1
Description:
Measurements are suspect due to evidence that the factory calibration for volumetric water 
content (VWC) measurements are insufficient for conditions at the site. Specifically for 
periods in which values are 9999999, indicates that the factory calibration of bulk 
permittivity to calculate VWC is not applicable. Soil bulk permittivity was not ingested for 
this period, so no back analysis is feasible.
Measurements:nsaamcC1.b1:
  • Average volumetric water content sensor 4(vwc_4)
  • Average volumetric water content sensor 3(vwc_3)
  • Average volumetric water content sensor 11(vwc_11)
  • Average volumetric water content sensor 2(vwc_2)
  • Average volumetric water content sensor 10(vwc_10)
  • Average volumetric water content sensor 1(vwc_1)
  • Average volumetric water content sensor 9(vwc_9)
  • Average volumetric water content sensor 7(vwc_7)
  • Average volumetric water content sensor 8(vwc_8)
  • Average volumetric water content sensor 5(vwc_5)
  • Average volumetric water content sensor 6(vwc_6)


Back To Table of Contents



END OF DATA