Data Quality Reports for Session: 184497 User: silberi1 Completed: 03/22/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160913.3AWR/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attributeawrceilS1.b1Does not affect quality


DQRID : D160913.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/03/2015210701/17/20162359Does not affect quality
Subject:
AWR/CEIL/S1 - Reprocess: Incorrect averaging_inverval global attribute
DataStreams:awrceilS1.b1
Description:
When the ceilometer software was changed from CT-View to CL-View, several firmware 
configuration changes were made to the model CT25K for consistency with the CL31 at other sites, 
including the data acquisition interval which was changed from 15 to 16 seconds. 
However, the global attribute "averaging_interval" in the ceil datastream was not changed.
Measurements:awrceilS1.b1:
  • base_time


Back To Table of Contents



END OF DATA