DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D140318.1 | TMP/MWR3C/M1 - Calibration not updated | tmpmwr3cM1.b1 | Suspect |
D150210.2 | TMP/MWR3C/M1 - Reprocess: Incorrect calibration | tmpmwr3cM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/10/2014 | 0000 | 02/13/2014 | 1400 | Suspect |
Subject: | TMP/MWR3C/M1 - Calibration not updated |
DataStreams: | tmpmwr3cM1.b1 |
Description: | The radiometer didn't update the calibration because the relative humidity was too high. The relative humidity threshold was increased and the radiometer updated its calibration. |
Measurements: | tmpmwr3cM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/23/2014 | 1300 | 02/03/2014 | 1419 | Incorrect |
Subject: | TMP/MWR3C/M1 - Reprocess: Incorrect calibration |
DataStreams: | tmpmwr3cM1.b1 |
Description: | The instrument is not calibrated data need to be reprocessed |
Measurements: | tmpmwr3cM1.b1:
|