DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D100224.6 | GRW/MWR/M1 - Reprocessed: Biased LWP retrievals-Updated retrieval coefficients | grwmwrlosM1.b1 | Does not affect quality |
D100402.3 | GRW/WACR/M1 - Coherence Filter Contamination | grwwacrM1.b1 | Suspect |
D100729.5 | GRW/WACR/M1 - Reprocessed: Reflectivity and Height Offsets | grwwacrM1.b1 | Does not affect quality |
D120301.1 | GRW/WACR/M1 - Transmit power drop outs | grwwacrM1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/01/2009 | 0000 | 03/31/2010 | 1700 | Does not affect quality |
Subject: | GRW/MWR/M1 - Reprocessed: Biased LWP retrievals-Updated retrieval coefficients |
DataStreams: | grwmwrlosM1.b1 |
Description: | Retrieval coefficients were updated to screen for some bad soundings. Data were reprocessed 20100507. |
Measurements: | grwmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/09/2009 | 1600 | 04/02/2010 | 1200 | Suspect |
Subject: | GRW/WACR/M1 - Coherence Filter Contamination |
DataStreams: | grwwacrM1.b1 |
Description: | The coherence filter (software) that was added to the WACR under BCR-01500 was causing occasional contamination to WACR data. This was exhibited in the first 2000 m of range by streaking in the data over a period of time. It didn't occur often and when it did, it is quite obvious looking at the plots. On 4/2/2010 at 12:00 UTC, the coherence filter was modified to preclude working on signals with a SNR greater than 60 dB. The vendor thinks this will solve the problem. |
Measurements: | grwwacrM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/14/2009 | 0000 | 01/06/2011 | 0800 | Does not affect quality |
Subject: | GRW/WACR/M1 - Reprocessed: Reflectivity and Height Offsets |
DataStreams: | grwwacrM1.b1 |
Description: | The instrument vendor informed us of a range (height) offset of 13 m and provided updated reflectivity calibration. All values reported for 'Reflectivity' were approximately 3 dB too low. In addition, the Reflectivity in the first and fourth range gates (starting from the surface) needed to be further corrected by an additional 2.7 and -0.9 dB, respectively. The 'height' field values were 13.33 m too low. The data were reprocessed to correct these problems. The grwwacrM1.b1 data were reprocessed and archived in October 2010. The grwwacrspeccmaskcopolM1.a0 and grwwacrspeccmaskxpolM1.a0 data were reprocessed and archived in January 2012. |
Measurements: | grwwacrM1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/27/2009 | 1627 | 10/19/2010 | 1237 | Incorrect |
Subject: | GRW/WACR/M1 - Transmit power drop outs |
DataStreams: | grwwacrM1.b1 |
Description: | At approximately 16:27 GMT on 20090727, txpower began intermittently dropping out and reporting slightly negative values. This issue continued intermittently through 20090807 when it stopped occurring. The problem resumed 20091006 and continued to occur on a daily basis until the monitor detector was replaced 20101019. The transmit power as measured by the detector after replacement was approximately 1.3 dB less than when it failed. Reflectivity measurements have a +0/-1.3 dBZ of additional uncertainty for this period of time. |
Measurements: | grwwacrM1.b1:
|