Data Quality Reports for Session: 184670 User: yangj4 Completed: 03/29/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D140821.4NSA/MERGESONDE/C1 - Reprocess: Suspect rh_scaled valuesnsamergesonde1maceC1.c1Suspect


DQRID : D140821.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/01/2002000012/31/20102359Suspect
Subject:
NSA/MERGESONDE/C1 - Reprocess: Suspect rh_scaled values
DataStreams:nsamergesonde1maceC1.c1
Description:
MERGESONDE sometimes uses bad or questionable precipitable water vapor data to scale the 
relative humidity profiles provided by the rh_scaled field. In addition, this bad input 
data is not properly indicated by the vapor_source or qc_rh_scaled fields. This problem 
appears to typically occur when there is moisture impacting the microwave radiometer causing 
large spikes in brightness temperatures. During these periods the rh_scaled values are 
often saturated at 101% for much of the profile.
Measurements:nsamergesonde1maceC1.c1:
  • vapor_source
  • rh_scaled


Back To Table of Contents