Data Quality Reports for Session: 184674 User: akinolao1 Completed: 03/29/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D100729.4HFE/WACR/M1 - Reprocessed: Reflectivity and Height OffsetshfewacrM1.b1Does not affect quality


DQRID : D100729.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/15/2008000012/15/20082359Does not affect quality
Subject:
HFE/WACR/M1 - Reprocessed: Reflectivity and Height Offsets
DataStreams:hfewacrM1.b1
Description:
The instrument vendor informed us of a range (height) offset of 13 m and provided updated 
reflectivity calibration.  All values reported for 'Reflectivity' were approximately 3 dB 
too low.   In addition, the Reflectivity in the first and fourth range gates (starting 
from the surface) needed to be further corrected by an additional 2.7 and -0.9 dB, 
respectively. The 'height' field values were 13.33 m too low.  The data were reprocessed to 
correct these problems.

The hfewacrM1.b1 data were reprocessed and archived in October 2010.  The
hfewacrspeccmaskcopolM1.a0 and hfewacrspeccmaskxpolM1.a0 data were reprocessed and
archived in January and July 2012.
Measurements:hfewacrM1.b1:
  • heights
  • Reflectivity(Reflectivity)


Back To Table of Contents



END OF DATA