Data Quality Reports for Session: 185012 User: weltyj1 Completed: 04/12/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D051031.3SGP/ECOR/E14 - CO2/H2O Incorrectsgp30ecorE14.b1Incorrect
D051216.2SGP/ECOR/E14 - Flux Measurements Incorrectsgp30ecorE14.b1Incorrect
D070223.1SGP/ECOR/E14 - Fluxes and Most Measurements Incorrectsgp30ecorE14.b1Incorrect
D070514.2SGP/ECOR/E14 - IRGA Temperature Incorrect or Missingsgp30ecorE14.b1Incorrect
D080604.2SGP/ECOR/E14 - Sensor failuresgp30ecorE14.b1Suspect


DQRID : D051031.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/25/2005000008/29/20050230Incorrect
Subject:
SGP/ECOR/E14 - CO2/H2O Incorrect
DataStreams:sgp30ecorE14.b1
Description:
The LI-7500 CO2/H2O sensor stopped working after a rainstorm.  CO2 and H2O data are 
incorrect during the stated period.
Measurements:sgp30ecorE14.b1:
  • n_bad_irga
  • skew_q
  • covariance tq(cvar_tq)
  • rotated covariance wc(cvar_rot_wc)
  • rotated covariance vq(cvar_rot_vq)
  • covariance vq(cvar_vq)
  • n_spk_c
  • temp_irga
  • Latent heat flux(lv_e)
  • kurt_c
  • covariance uc(cvar_uc)
  • Mean CO2 density(mean_c)
  • covariance wc(cvar_wc)
  • Variance of q(var_q)
  • rotated covariance vc(cvar_rot_vc)
  • rotated covariance wq(cvar_rot_wq)
  • n_good_q
  • real_lv
  • Mean H2O density(mean_q)
  • kurt_q
  • covariance uq(cvar_uq)
  • mean_spk_c
  • Variance of c(var_c)
  • covaraince wq(cvar_wq)
  • n_good_c
  • rotated covariance uq(cvar_rot_uq)
  • n_bad_c
  • CO2 flux(fc)
  • n_bad_irga_light
  • covariance tc(cvar_tc)
  • covaraince vc(cvar_vc)
  • covariance qc(cvar_qc)
  • skew_c
  • rotated covariance uc(cvar_rot_uc)
  • n_spk_q
  • n_bad_q
  • mean_spk_q


Back To Table of Contents

DQRID : D051216.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/07/2006193007/14/20061500Incorrect
11/04/2005133011/17/20051630Incorrect
01/24/2006150002/13/20061930Incorrect
05/05/2006210006/13/20061400Incorrect
12/01/2005193012/13/20051630Incorrect
Subject:
SGP/ECOR/E14 - Flux Measurements Incorrect
DataStreams:sgp30ecorE14.b1
Description:
The ECOR datastream was corrupted; this happens when the ECOR program produces a 
datastream of the wrong length (i.e., leaves one or more measurements out).

Fluxes and some other measurements are incorrect.

Reprocessing will not correct this problem; the raw data are corrupted.
Measurements:sgp30ecorE14.b1:
  • Mean H2O density(mean_q)
  • Corrected sensible heat flux(h)
  • atm_pres
  • Mean CO2 density(mean_c)
  • temp_irga
  • CO2 flux(fc)
  • Latent heat flux(lv_e)


Back To Table of Contents

DQRID : D070223.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/14/2007050002/02/20071600Incorrect
Subject:
SGP/ECOR/E14 - Fluxes and Most Measurements Incorrect
DataStreams:sgp30ecorE14.b1
Description:
The ECOR system was having major problems during this period and most measurements were 
incorrect most of the time.
Measurements:sgp30ecorE14.b1:
  • n_bad_irga
  • rotated covariance vq(cvar_rot_vq)
  • rotated covariance vw(cvar_rot_vw)
  • temp_irga
  • rotated covariance vt(cvar_rot_vt)
  • rho
  • Mean CO2 density(mean_c)
  • rotated covariance vc(cvar_rot_vc)
  • rotated covariance uw(cvar_rot_uw)
  • Altitude above mean sea level(alt)
  • rotated covariance uq(cvar_rot_uq)
  • standard deviation of wind direction(std_wind_dir)
  • rotated covariance ut(cvar_rot_ut)
  • rotated covariance uv(cvar_rot_uv)
  • CO2 flux(fc)
  • Variance of v(var_rot_v)
  • Variance of rotated u(var_rot_u)
  • Variance of rotated w(var_rot_w)
  • covariance tc(cvar_tc)
  • mean_cooler
  • skew_c
  • East longitude(lon)
  • rotated covariance uc(cvar_rot_uc)
  • n_bad_son_ic
  • Water vapor mixing ratio observed by the Raman lidar(mr)
  • n_good_w
  • variance of variable v(var_v)
  • variance of variable u(var_u)
  • n_good_v
  • covariance tq(cvar_tq)
  • n_good_u
  • variance of variable w(var_w)
  • n_good_t
  • elev
  • Latent heat flux(lv_e)
  • base_time
  • covariance uc(cvar_uc)
  • Variance of q(var_q)
  • variance of variable t(var_t)
  • n_good_q
  • wind_spd
  • covariance ut(cvar_ut)
  • k
  • covariance uv(cvar_uv)
  • Corrected sensible heat flux(h)
  • lv
  • mean_rot_w
  • covariance uq(cvar_uq)
  • Mean horizontal wind speed(mean_rot_u)
  • mean_rot_v
  • time_offset
  • mean_spk_c
  • Variance of c(var_c)
  • n_good_c
  • n_bad_irga_light
  • wind_dir
  • mean_spk_v
  • mean_spk_w
  • covaraince vc(cvar_vc)
  • mean_spk_t
  • mean_spk_u
  • covariance uw(cvar_uw)
  • mean_spk_q
  • rotated covariance wc(cvar_rot_wc)
  • covariance vq(cvar_vq)
  • n_spk_c
  • covariance vw(cvar_vw)
  • covariance vt(cvar_vt)
  • kurt_c
  • covariance wc(cvar_wc)
  • rotated covariance wt(cvar_rot_wt)
  • rotated covariance wq(cvar_rot_wq)
  • real_lv
  • kurt_u
  • kurt_t
  • Pressure, condensation, at altitude, RUC model output(cp)
  • kurt_q
  • covaraince wq(cvar_wq)
  • friction velocity(ustar)
  • covariance wt(cvar_wt)
  • real_cp
  • n_spk_w
  • n_spk_t
  • n_spk_u
  • n_spk_v
  • n_spk_q
  • kurt_v
  • real_mr
  • kurt_w
  • skew_u
  • skew_t
  • skew_w
  • skew_v
  • skew_q
  • standard deviation of wind elevation angle(std_elev)
  • real_rho
  • atm_pres
  • time
  • Mean H2O density(mean_q)
  • Mean sonic temp(mean_t)
  • mean_u
  • mean_v
  • phi
  • mean_w
  • n_bad_c
  • covariance qc(cvar_qc)
  • n_bad_w
  • n_bad_v
  • n_bad_u
  • n_bad_t
  • n_bad_q
  • North latitude(lat)


Back To Table of Contents

DQRID : D070514.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/09/2007130005/19/20071700Incorrect
Subject:
SGP/ECOR/E14 - IRGA Temperature Incorrect or Missing
DataStreams:sgp30ecorE14.b1
Description:
The temp_irga was too large 05/09/07 1300 GMT - 05/10/07 2000 GMT.

The temp_irga was flat lined at 49.96 C 05/10/07 2030 GMT - 05/17/07 2000 GMT.  

The temp_irga was missing 05/17/07 2030 GMT - 05/19/07 1700 GMT.
Measurements:sgp30ecorE14.b1:
  • temp_irga


Back To Table of Contents

DQRID : D080604.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/02/2008133006/05/20081930Suspect
Subject:
SGP/ECOR/E14 - Sensor failure
DataStreams:sgp30ecorE14.b1
Description:
The water vapor output of the LI-7500 was too low, indicating that water vapor leaked into 
the sensor and that the chemicals are spent or there was an electronic failure.

Latent heat flux was not normally affected by this problem.
Measurements:sgp30ecorE14.b1:
  • Mean H2O density(mean_q)


Back To Table of Contents



END OF DATA