Data Quality Reports for Session: 185068 User: dturner Completed: 04/17/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160520.12SGP/DL/DLPROF/C1 - Reprocess: Incorrect serial numbersgpdlfptC1.b1, sgpdlprofwind4newsC1.c1Does not affect quality
D160520.13SGP/DL/DLPROF/E32 - Reprocessed and Corrected: Incorrect serial numbersgpdlfptE32.b1Does not affect quality
D160520.14SGP/DL/DLPROF/E37 - Reprocessed and Corrected: Incorrect serial numbersgpdlfptE37.b1Does not affect quality
D160617.10SGP/DL/C1 - Power supply issuesgpdlfptC1.b1Missing
D160708.11SGP/MWR3C/E39 - Condensation on lenssgpmwr3cE39.b1Incorrect
D160708.4SGP/MWR3C/E37 - Uncertain IRT datasgpmwr3cE37.b1Suspect
D160708.5SGP/MWR3C/E37 - Instrument malfunctionsgpmwr3cE37.b1Incorrect
D161021.7SGP/MET/E13 - False corrected precipitation valuessgpmetE13.b1Incorrect
D161109.3SGP/MWR3C/C1 - Reprocess: Incorrect temperature coefficientsgpmwr3cC1.b1Suspect
D170322.11SGP/MET/E37 - Temperature/humidity sensor not calibrated annuallysgpmetE37.b1Does not affect quality


DQRID : D160520.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/02/2016180005/19/20162200Does not affect quality
Subject:
SGP/DL/DLPROF/C1 - Reprocess: Incorrect serial number
DataStreams:sgpdlfptC1.b1, sgpdlprofwind4newsC1.c1
Description:
The instrument serial number that appears in the netcdf files global attributes is 
incorrect. The correct serial number is 0116-107.
Measurements:sgpdlprofwind4newsC1.c1:
  • base_time

sgpdlfptC1.b1:
  • base_time


Back To Table of Contents

DQRID : D160520.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/03/2016180005/17/20161700Does not affect quality
Subject:
SGP/DL/DLPROF/E32 - Reprocessed and Corrected: Incorrect serial number
DataStreams:sgpdlfptE32.b1
Description:
DATA REPROCESSED and CORRECTED: 

The instrument serial number that appears in the netcdf files global attributes was 
incorrect. The corrected serial number is 0116-108.

The data were archived March 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::dl/fsite::sgp/ffac::sgp.E32 


ORIGINAL DQR CATEGORY: Transparent( Does not affect quality)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-03

REPROCESS IDENTIFICATION NUMBER: RID 1466

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------

The instrument serial number that appears in the netcdf files global attributes is 
incorrect. The correct serial number is 0116-108.
Measurements:sgpdlfptE32.b1:
  • base_time


Back To Table of Contents

DQRID : D160520.14
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/03/2016150005/19/20162200Does not affect quality
Subject:
SGP/DL/DLPROF/E37 - Reprocessed and Corrected: Incorrect serial number
DataStreams:sgpdlfptE37.b1
Description:
DATA REPROCESSED and CORRECTED: 

The instrument serial number that appears in the netcdf files global attributes is 
incorrect. The corrected serial number is 0116-109.

The data were archived in March 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::dl/fsite::sgp/ffac::sgp.E37 


ORIGINAL DQR CATEGORY: Transparent( Does not affect quality)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-03

REPROCESS IDENTIFICATION NUMBER: RID 1467

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------

The instrument serial number that appears in the netcdf files global attributes is 
incorrect. The correct serial number is 0116-109.
Measurements:sgpdlfptE37.b1:
  • base_time


Back To Table of Contents

DQRID : D160617.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/27/2016202005/31/20161627Missing
Subject:
SGP/DL/C1 - Power supply issue
DataStreams:sgpdlfptC1.b1
Description:
The supply voltage from the external power supply was running low due to overheating. As a 
result, the instrument was intentionally shutdown for the long Memorial Day weekend.
Measurements:sgpdlfptC1.b1:
  • time_offset
  • Altitude above mean sea level(alt)
  • Attenuated backscatter(attenuated_backscatter)
  • qc_time
  • base_time
  • North latitude(lat)
  • qc_radial_velocity
  • azimuth
  • range
  • Radial velocity(radial_velocity)
  • intensity
  • East longitude(lon)
  • time
  • elevation


Back To Table of Contents

DQRID : D160708.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/07/2016100006/07/20161400Incorrect
06/19/2016070006/19/20161400Incorrect
06/04/2016030006/04/20161400Incorrect
06/16/2016100006/16/20161400Incorrect
05/31/2016050005/31/20161400Incorrect
04/14/2016100004/14/20161400Incorrect
05/10/2016060005/10/20161400Incorrect
06/08/2016070006/08/20161400Incorrect
06/06/2016100006/06/20161400Incorrect
04/21/2016100004/21/20161400Incorrect
07/04/2016100007/04/20161300Incorrect
06/14/2016070006/14/20161400Incorrect
06/12/2016050006/12/20161400Incorrect
05/27/2016100005/27/20161400Incorrect
05/28/2016080005/28/20161400Incorrect
06/20/2016070006/20/20161400Incorrect
06/11/2016060006/11/20161400Incorrect
05/26/2016100005/26/20161400Incorrect
06/15/2016070006/15/20161400Incorrect
04/15/2016100004/15/20161400Incorrect
06/21/2016090006/21/20161300Incorrect
06/13/2016070006/13/20161400Incorrect
06/03/2016030006/03/20161400Incorrect
05/29/2016050005/29/20161400Incorrect
07/02/2016110007/02/20161300Incorrect
Subject:
SGP/MWR3C/E39 - Condensation on lens
DataStreams:sgpmwr3cE39.b1
Description:
The dewblower was malfunctioning and there was condensation on the lens when the RH was 
higher that ~80%.
Measurements:sgpmwr3cE39.b1:
  • 89.0 GHz sky brightness temperature(tbsky89)
  • 30.0 GHz sky brightness temperature(tbsky30)
  • 23.8 GHz sky brightness temperature(tbsky23)
  • Liquid water path best-estimate value, hourly mean(lwp)
  • Column precip water vapor(pwv)


Back To Table of Contents

DQRID : D160708.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/28/2016193506/07/20161550Suspect
Subject:
SGP/MWR3C/E37 - Uncertain IRT data
DataStreams:sgpmwr3cE37.b1
Description:
The IRT data appear to disagree with the co-located MET IRT.
Measurements:sgpmwr3cE37.b1:
  • infrared_temperature


Back To Table of Contents

DQRID : D160708.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/30/2016160006/17/20161508Incorrect
Subject:
SGP/MWR3C/E37 - Instrument malfunction
DataStreams:sgpmwr3cE37.b1
Description:
The instrument malfunctioned and was pulled out and sent to the vendor
Measurements:sgpmwr3cE37.b1:
  • Column precip water vapor(pwv)
  • Liquid water path best-estimate value, hourly mean(lwp)
  • 23.8 GHz sky brightness temperature(tbsky23)
  • 30.0 GHz sky brightness temperature(tbsky30)


Back To Table of Contents

DQRID : D161021.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/21/2016185504/21/20161905Incorrect
03/11/2016190703/11/20161909Incorrect
08/25/2016191508/25/20161916Incorrect
08/11/2016141508/11/20161500Incorrect
05/20/2016154505/20/20161547Incorrect
06/30/2016190706/30/20161912Incorrect
06/17/2016135706/17/20161359Incorrect
02/26/2016154502/26/20161547Incorrect
06/03/2016162106/03/20161622Incorrect
09/23/2016152009/23/20161521Incorrect
04/07/2016155404/07/20161556Incorrect
09/08/2016142709/08/20161429Incorrect
07/12/2016134107/12/20161343Incorrect
01/29/2016154001/29/20161541Incorrect
05/05/2016183005/05/20161831Incorrect
02/11/2016192702/11/20161929Incorrect
03/25/2016152203/25/20161524Incorrect
Subject:
SGP/MET/E13 - False corrected precipitation values
DataStreams:sgpmetE13.b1
Description:
A program bug allowed for maintenance tips to be recorded as precipitation in the 
tbrg_precip_total_corr variable.
Measurements:sgpmetE13.b1:
  • Corrected total precipitation measured by TBRG(tbrg_precip_total_corr)


Back To Table of Contents

DQRID : D161109.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/01/2016000011/09/20161728Suspect
Subject:
SGP/MWR3C/C1 - Reprocess: Incorrect temperature coefficient
DataStreams:sgpmwr3cC1.b1
Description:
The 30 GHz temperature correction coefficient has slightly changed. This results in a 
diurnal temperature dependence of the brightness temperature and retrievals.
Measurements:sgpmwr3cC1.b1:
  • 30.0 GHz sky brightness temperature(tbsky30)
  • Liquid water path best-estimate value, hourly mean(lwp)
  • Column precip water vapor(pwv)


Back To Table of Contents

DQRID : D170322.11
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/30/2011000009/12/20162359Does not affect quality
Subject:
SGP/MET/E37 - Temperature/humidity sensor not calibrated annually
DataStreams:sgpmetE37.b1
Description:
The temperature/humidity sensor was not calibrated annually during this time, but passed 
all 6-month field calibration checks. Comparison with a nearby Mesonet site suggests the 
relative humidity may be reading low, especially in high humidity situations. An exit 
calibration will characterize severity when the sensor is replaced.

In general, the relative humidity may have been subject to drift during this time. The 
manufacturer characterizes sensor drift at <1% / year.
Measurements:sgpmetE37.b1:
  • Temperature mean(temp_mean)
  • Vapor pressure mean, calculated(vapor_pressure_mean)
  • Humidity, relative, at 2-m height, 1-min avg(rh_mean)


Back To Table of Contents



END OF DATA