Data Quality Reports for Session: 185089 User: liuj3 Completed: 04/19/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D160119.12AWR/HTDMA/M1 - HT-DMA CPC zero countsawraoshtdmaM1.a1Incorrect
D160201.1AWR/HTDMA/M1 - Low instrument flowsawraoshtdmaM1.a1Incorrect
D160209.2AWR/HTDMA/M1 - No particle countsawraoshtdmaM1.a1Incorrect
D160209.3AWR/HTDMA/M1 - Intermittent zero countsawraoshtdmaM1.a1Incorrect
D160218.4AWR/HTDMA/M1 - Instrument reliability issuesawraoshtdmaM1.a1Incorrect
D161130.3AWR/HTDMA/M1 - Instrument flows not nominalawraoshtdmaM1.a1Does not affect quality
D161130.4AWR/HTDMA/M1 - Instrument connectivity and flow issuesawraoshtdmaM1.a1Incorrect
D161216.7AWR/HTDMA/M1 - Initial size selection 200 nmawraoshtdmaM1.a1Suspect


DQRID : D160119.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/20/2015120001/16/20160000Incorrect
Subject:
AWR/HTDMA/M1 - HT-DMA CPC zero counts
DataStreams:awraoshtdmaM1.a1
Description:
HT-DMA CPC butanol flooding or contamination. CPC was showing zero counts. Intermittent 
counts from troubleshooting. No valid data.
Measurements:awraoshtdmaM1.a1:
  • Total concentration(total_concentration)
  • scan_max_diameter_setting
  • bin_width
  • bin_center
  • num_bins
  • bin
  • dry_diameter_setting
  • dry_rh_setting
  • Concentration of aerosol(aerosol_concentration)
  • scan_min_diameter_setting


Back To Table of Contents

DQRID : D160201.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/29/2016120001/30/20160800Incorrect
Subject:
AWR/HTDMA/M1 - Low instrument flows
DataStreams:awraoshtdmaM1.a1
Description:
External vacuum was lost, affecting instrument flows. No valid data from the time period.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • bin
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D160209.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/18/2015014011/20/20152300Incorrect
Subject:
AWR/HTDMA/M1 - No particle counts
DataStreams:awraoshtdmaM1.a1
Description:
No particle counts after the instrument was turned on. Probably no butanol flow to the 
instrument or possibly a disconnected pump. No meaningful data saved.
Measurements:awraoshtdmaM1.a1:
  • Total concentration(total_concentration)
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D160209.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/30/2015170012/02/20151900Incorrect
Subject:
AWR/HTDMA/M1 - Intermittent zero counts
DataStreams:awraoshtdmaM1.a1
Description:
Particle counts intermittently going to zero. Probably blocked butanol flow to the 
instrument or possibly internal air flow instability.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D160218.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/31/2016030011/08/20162100Incorrect
Subject:
AWR/HTDMA/M1 - Instrument reliability issues
DataStreams:awraoshtdmaM1.a1
Description:
HT-DMA particle counts intermittently dropping to zero due to internal issues of the 
HT-DMA MCPC. Instrument periodically sampling cabin air for troubleshooting or switched off 
while waiting for spare parts to arrive.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • bin
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D161130.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/08/2016210001/02/20171416Does not affect quality
Subject:
AWR/HTDMA/M1 - Instrument flows not nominal
DataStreams:awraoshtdmaM1.a1
Description:
SEMS part of the HT-DMA has a DMA sheath flow imbalance. Growth factor measurement not 
affected.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • bin
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D161130.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/20/2016020011/21/20161700Incorrect
11/16/2016133011/17/20161600Incorrect
11/22/2016170011/24/20161600Incorrect
Subject:
AWR/HTDMA/M1 - Instrument connectivity and flow issues
DataStreams:awraoshtdmaM1.a1
Description:
HSEMS intermittently losing connection with SEMS and not setting the classifier diameter. 
MCPC not drawing in butanol due to internal pressure imbalance. Data incorrect.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • bin
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents

DQRID : D161216.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/13/2016010012/13/20161445Suspect
Subject:
AWR/HTDMA/M1 - Initial size selection 200 nm
DataStreams:awraoshtdmaM1.a1
Description:
Communication lost between the two major parts of the HT-DMA system. Initial size 
selection stuck at 200 nm.
Measurements:awraoshtdmaM1.a1:
  • scan_max_diameter_setting
  • Total concentration(total_concentration)
  • scan_time_setting
  • bin_width
  • num_bins
  • dry_diameter_setting
  • dry_rh_setting
  • sample_flow
  • humid_rh
  • bin_center
  • bin
  • atomizer
  • scan_type
  • scan_min_diameter_setting
  • Concentration of aerosol(aerosol_concentration)


Back To Table of Contents



END OF DATA