Data Quality Reports for Session: 185093 User: sherman Completed: 04/18/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D050321.2SGP/TSI/C1 - Instrument failuresgptsiC1.00Missing
D060405.3SGP/TSI/C1 - Reprocessing Notice (RID 253)sgptsiC1.00Does not affect quality


DQRID : D050321.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/17/2005192504/05/20051630Missing
Subject:
SGP/TSI/C1 - Instrument failure
DataStreams:sgptsiC1.00
Description:
The TSI-880 experienced a power and/or computer failure. It was replaced with a spare 
TSI-440 and data collection resumed 4/5/05.  Note: this is a temporary replacement.
Measurements:sgptsiC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents

DQRID : D060405.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/30/2005000004/04/20062019Does not affect quality
Subject:
SGP/TSI/C1 - Reprocessing Notice (RID 253)
DataStreams:sgptsiC1.00
Description:
See below for details of reprocessing that was completed on 20150330.  Original 
DQR category: 5 (Transparent - Does not affect quality).

-----------------------------------------------------------

When the TSI-660 was installed and replaced the temporary TSI-440, I forgot to change the 
serial number in the tsi.properties file. These data could be reprocessed to change 
attribute serial_number from "440" to "106".
Measurements:sgptsiC1.00:
  • Raw data stream - documentation not supported


Back To Table of Contents



END OF DATA