D150818.1MAO/SKYRAD/M1 - Reprocess: Incorrect calibration coefficientsmaoskyrad60sM1.b1Incorrect


DQRID : D150818.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/05/2015173008/13/20151900Incorrect
Subject:
MAO/SKYRAD/M1 - Reprocess: Incorrect calibration coefficients
DataStreams:maoskyrad60sM1.b1
Description:
During the DQR time period, the CR3000 program had correct serial numbers and calibration 
coefficients for the PIR1-DIR and PIR2-DIR datalogger channels:  PIR1 SN34303, 
calibration coefficient K1= 0.247520; PIR2 SN34306, calibration coefficient K1=0.230410.
PIR SN34306 (on sun tracker) was used with K1= 0.247520 on the PIR1-DIR datalogger 
channel; K1=0.230410 should be used for reprocessing to provide correct data for PIR1-DIR 
parameters.
PIR SN34303 (on sun tracker) was used with K1= 0.230410 on the PIR2-DIR datalogger 
channel; K1=0.247520 should be used for reprocessing to provide correct data for PIR2-DIR 
parameters.
Measurements:maoskyrad60sM1.b1:
  • down_long_hemisp2_max
  • inst_down_long_hemisp1_dome_temp
  • down_long_hemisp1_min
  • inst_down_long_hemisp2_case_temp
  • inst_down_long_hemisp2_tp
  • down_long_hemisp1_std
  • down_long_hemisp1_max
  • inst_down_long_hemisp1_tp
  • inst_down_long_hemisp1_case_temp
  • down_long_hemisp2_min
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp2)
  • inst_down_long_hemisp2_dome_temp
  • down_long_hemisp2_std


Back To Table of Contents