Data Quality Reports for Session: 185129 User: chenx3 Completed: 04/20/2017


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected Data Quality Metric
D060828.2SGP/IRT/C1 - UPS failingsgpirt10mC1.b1Suspect
D120113.6SGP/IRT/C1 - Noisy 10m IRT datasgpirt10mC1.b1Incorrect
D121022.1SGP/IRT/C1 - Measurement rangesgpirt10mC1.b1Suspect
D130211.1SGP/IRT/C1 - 10m IRT ground surface temperaturesgpirt10mC1.b1Does not affect quality
D130806.2SGP/IRT/C1 - Noisy datasgpirt10mC1.b1Suspect


DQRID : D060828.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/26/2006085309/18/20061530Suspect
Subject:
SGP/IRT/C1 - UPS failing
DataStreams:sgpirt10mC1.b1
Description:
The IRT data were intermittently bad during this period due to a failing UPS. The problem 
was corrected by replacing the UPS.
Measurements:sgpirt10mC1.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • internal_voltage
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D120113.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/03/2012153001/04/20122030Incorrect
Subject:
SGP/IRT/C1 - Noisy 10m IRT data
DataStreams:sgpirt10mC1.b1
Description:
The 10M IRT surface temperatures were very noisy because the connection to the data logger 
had become loose.
Measurements:sgpirt10mC1.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • internal_voltage
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)
  • logger_temp


Back To Table of Contents

DQRID : D121022.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/25/2012180008/04/20122059Suspect
Subject:
SGP/IRT/C1 - Measurement range
DataStreams:sgpirt10mC1.b1
Description:
The ground surface temperature exceeded the maximum measurement of 50C on some days. The 
maximum was increased to 60C for IRT10m on 10/1/2012 at 15:25 and for IRT25m on 10/9/2012 
at 21:45.
Measurements:sgpirt10mC1.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)


Back To Table of Contents

DQRID : D130211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/11/2012075002/08/20131525Does not affect quality
Subject:
SGP/IRT/C1 - 10m IRT ground surface temperature
DataStreams:sgpirt10mC1.b1
Description:
The 10m IRT was measuring lower ground surface temperature than the 25m IRT. The lens was 
cleaned on 12/19/2012 and replaced on 2/8/2013 but the difference in temperature appears 
to be due to the seasonal difference in surface cover - tall grass vesus bare soil.
Measurements:sgpirt10mC1.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)


Back To Table of Contents

DQRID : D130806.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/05/2011160307/14/20110303Suspect
08/03/2011224208/12/20110627Suspect
07/26/2011202107/26/20112310Suspect
Subject:
SGP/IRT/C1 - Noisy data
DataStreams:sgpirt10mC1.b1
Description:
The power supplied to the data logger was very "noisy" causing the surface temperature 
measurements to also be "noisy".
Measurements:sgpirt10mC1.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • internal_voltage
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • sfc_ir_temp_std
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)


Back To Table of Contents



END OF DATA