DQR ID | Subject | Data Streams Affected | Data Quality Metric |
---|---|---|---|
D120827.3 | SGP/ACSM/C1 - Water in inlet sample line | sgpaosacsmC1.b1 | Incorrect |
D121211.5 | SGP/ACSM/C1 - Pump failure | sgpaosacsmC1.b1 | Incorrect |
D130108.1 | SGP/ACSM/C1 - Pump failure | sgpaosacsmC1.b1 | Incorrect |
D130108.2 | SGP/ACSM/C1 - Sample flow restricted | sgpaosacsmC1.b1 | Suspect |
D130201.2 | SGP/ACSM/C1 - Missing data | sgpaosacsmC1.b1 | Missing |
D140417.1 | SGP/ACSM/C1 - Missing data | sgpaosacsmC1.b1 | Missing |
D141009.2 | SGP/ACSM/C1 - Reprocessing Notice (RID 1237) | sgpaosacsmC1.b1 | Does not affect quality |
D150410.1 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D150410.2 | SGP/ACSM/C1 - missing data | sgpaosacsmC1.b1 | Missing |
D160715.4 | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing | sgpaosacsmC1.b1 | Does not affect quality |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/14/2012 | 1500 | 07/31/2012 | 2359 | Incorrect |
06/03/2012 | 1800 | 06/12/2012 | 0700 | Incorrect |
Subject: | SGP/ACSM/C1 - Water in inlet sample line |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data became progressively more noisy causing large spikes in data. These spikes started occurring for longer periods of time during the day before the instrument went down completely on 7/12. The afternoon data (usually 15:00-24:00 UTC) was incorrect because of water condensation in the inlet line and aerosol collection efficiency change. The organic mass loading is around 0 when the condensed water was absorbing all the particles. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/13/2012 | 2200 | 12/05/2012 | 2210 | Incorrect |
Subject: | SGP/ACSM/C1 - Pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The carbon vane vacuum pump that pulls air through the instrument failed 11/13 following a power failure. The pump was replaced and returned to service at 2210 GMT on 12/05/2012. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/18/2012 | 1500 | 12/21/2012 | 2130 | Incorrect |
Subject: | SGP/ACSM/C1 - Pump failure |
DataStreams: | sgpaosacsmC1.b1 |
Description: | The carbon vane vacuum pump that pulls air through the instrument failed again and a section of tubing between the ACSM and pump was partially collapsed. The pump and tubing were replaced and returned to service at 2223 GMT on 12/18/2012. The ACSM was then tuned before resuming data collection at 2130 GMT on 12/21/2012. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/05/2012 | 2211 | 12/18/2012 | 1459 | Suspect |
Subject: | SGP/ACSM/C1 - Sample flow restricted |
DataStreams: | sgpaosacsmC1.b1 |
Description: | This period is between two failures of the carbon vane pump that pulls air through the instrument. During this time the tubing between the pump and ACSM collapsed and may have affected the sample flow rate. The collapsed tubing was replaced with a different type, one that is much stiffer and less prone to collapsing under vacuum. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2012 | 0900 | 11/13/2012 | 2200 | Missing |
Subject: | SGP/ACSM/C1 - Missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/01/2013 | 0000 | 07/03/2013 | 2359 | Missing |
06/01/2013 | 0000 | 06/13/2013 | 2359 | Missing |
02/27/2013 | 0000 | 02/27/2013 | 2359 | Missing |
11/22/2012 | 0000 | 12/05/2012 | 2359 | Missing |
07/01/2012 | 0000 | 07/30/2012 | 2117 | Missing |
10/26/2013 | 0000 | 10/31/2013 | 1418 | Missing |
05/01/2013 | 0000 | 05/24/2013 | 2359 | Missing |
Subject: | SGP/ACSM/C1 - Missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/01/2012 | 0000 | 10/09/2014 | 0000 | Does not affect quality |
Subject: | SGP/ACSM/C1 - Reprocessing Notice (RID 1237) |
DataStreams: | sgpaosacsmC1.b1 |
Description: | See below for details of reprocessing that was completed on 20150610. Original DQR category: 4 (Red, Incorrect). ----------------------------------------------------------- The data will be reprocessed with the manufacturer's software to correct errors in the calibration values, collection efficiencies, and instrument response. The data in the current form is incorrect and should not be used. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/29/2011 | 1307 | 08/02/2011 | 1416 | Missing |
08/08/2011 | 1012 | 08/09/2011 | 2154 | Missing |
12/31/2011 | 2216 | 12/31/2011 | 2320 | Missing |
11/23/2011 | 1839 | 11/25/2011 | 1529 | Missing |
10/18/2011 | 0906 | 10/18/2011 | 1042 | Missing |
12/30/2011 | 2023 | 12/30/2011 | 2126 | Missing |
12/31/2011 | 1938 | 12/31/2011 | 2042 | Missing |
07/14/2011 | 2250 | 07/18/2011 | 1945 | Missing |
12/27/2011 | 1755 | 12/27/2011 | 1859 | Missing |
03/01/2011 | 0000 | 03/01/2011 | 2359 | Missing |
10/09/2011 | 1256 | 10/09/2011 | 1947 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data for the specified periods are missing because if instrument problems. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/11/2012 | 2201 | 11/15/2012 | 0051 | Missing |
10/03/2012 | 0140 | 10/03/2012 | 1832 | Missing |
03/21/2012 | 1058 | 03/21/2012 | 1233 | Missing |
04/01/2012 | 0102 | 04/03/2012 | 0913 | Missing |
05/23/2012 | 1806 | 05/23/2012 | 1909 | Missing |
07/02/2012 | 1623 | 07/03/2012 | 2105 | Missing |
07/12/2012 | 1038 | 07/30/2012 | 2221 | Missing |
06/01/2012 | 0000 | 07/01/2012 | 0000 | Missing |
11/20/2012 | 1358 | 11/21/2012 | 1633 | Missing |
Subject: | SGP/ACSM/C1 - missing data |
DataStreams: | sgpaosacsmC1.b1 |
Description: | Data for the periods specified is missing because of instrument problems. |
Measurements: | sgpaosacsmC1.b1:
|
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/01/2011 | 0012 | 09/30/2014 | 2359 | Does not affect quality |
Subject: | SGP/ACSM/NEPHELOMETER/TDMA/C1 - Data Quality Evaluation Ongoing |
DataStreams: | sgpaosacsmC1.b1 |
Description: | A closure analysis is currently being perform on the ACSM, TDMA, and wet and dry NOAA NEPHELOMETERS at SGP C1. If problems are found during this analysis, they will be promptly documented through the DQR system. The results of this analysis will be made available once completed. Please contact the respective instrument mentors with any questions regarding these data over the specified period. |
Measurements: | sgpaosacsmC1.b1:
|